Table of Contents Table of Contents
Previous Page  94 / 1096 Next Page
Information
Show Menu
Previous Page 94 / 1096 Next Page
Page Background

S81

ESTRO 36

_______________________________________________________________________________________________

Conclusion

We have shown that an

a priori

scatter correction

algorithm for CB projections improves CBCT image quality

on both photon- and proton therapy gantries, potentially

opening for CBCT-based image/dose-guided proton

therapy.

OC-0159 Dual energy CBCT increases soft tissue CNR

ratio and image quality compared to standard CBCT in

IGRT

M. Skaarup

1

, D. Kovacs

1

, M.C. Aznar

1

, J.P. Bangsgaard

1

,

J.S. Rydhög

1

, L.A. Rechner

1

1

The Finsen Center - Rigshospitalet, Clinic of Oncology,

Copenhagen, Denmark

Purpose or Objective

We investigate a method for enhancing soft tissue contrast

to noise ratio (CNR) and clinical image quality of cone-

beam computed tomography (CBCT) by using a dual energy

CBCT protocol.

Material and Methods

Nine patients were scanned using a standard CBCT

protocol of either 100 or 125 kVp and a DE-CBCT protocol

of two separate scans of 80 and 140 kVp respectively.

Other scan parameters were identical and total radiation

dose was kept at a similar level for both protocols. Virtual

monochromatic dual energy (VMDE) images were

reconstructed using a linear mix of the 80 and 140 kVp

scan.

The weight, with which the two images were combined,

was calculated based on known attenuation coefficients of

two basis materials at a specific monochromatic energy. A

linear combination of these can be used to express the

attenuation coefficients of the 80 and 140 kVp scan at that

same monochromatic energy. To find the optimal virtual

reconstruction energy for soft tissue imaging, multiple

reconstructions were done for energies in the range 40-

180 keV.

CNR measurements were performed on both standard

CBCT and VMDE images for a number of different tissue

combinations, e.g. contrast between tumour-fat, tumour-

surrounding tissue, muscle-fat, rectum-surrounding

tissue, parotid-fat, seminal vesicle-surrounding tissue and

lung-heart (see figure 1 for an example). In addition, 5

experienced observers conducted a blinded ranking

between VMDE images (reconstructed at 55, 65, 75 and

100 keV) and the standard CBCT images, i.e. five image

series per patient. For each combination of image series

the observers were asked to compare the images side-by-

side, focusing on soft tissue image quality as well as

presence of metal artefacts. To eliminate left-right bias,

each combination was shown twice.

Results

VMDE images reconstructed at energies in the range 60 to

70 keV showed improved CNR for all soft tissue regions

when compared to the standard CBCT. On average, the

reconstruction energy corresponding to the maximum CNR

improvement is 65.5 ± 2.4 keV. The increase in maximum

CNR varied from 29% to 78%.

The clinical observer comparison gave a series of rankings

for each image series for each patient (see table 1). Using

signed rank Wilcoxon comparison test, the observers found

the VMDE images at 65 keV preferable to the standard

CBCT image. The p-value was found to be < 0.01, where p

< 0.05 is considered significant. An estimate of inter-

observer variability test was done with Fleiss’ kappa and

found to be moderate with a κ-value of 0.47, where values

above 0.4 is considered acceptable and 1 is perfect

agreement. Occasionally, an observer ranked the 75 keV

reconstruction as the most preferable image while the

overall preferred image was the 65 keV reconstruction.

Except in the case of patient one where the standard CBCT

image was ranked the highest of all.

Conclusion

VMDE images can increase soft tissue contrast and improve

clinical image quality for image-guided radiotherapy

compared to the standard CBCT protocol.

OC-0160 Radiomics Features Harmonization for CT

and CBCT in Rectal Cancer

R. Luo

1

, J. Wang

1

, H. Zhong

1

, J. Gan

1

, P. Hu

1

, L. Shen

1

,

W. Hu

1

, Z. Zhang

1

1

Fudan University Cancer Hospital, Radiation Oncology,

Shanghai, China

Purpose or Objective

Inter-scanner variability can lead to degradation of

radiomics

model

quality.

Therefore,

feature

harmonization is necessary for consistent findings in

radiomics studies, especially for multi-institution studies.

The purpose of this study is

to establish harmonization