Table of Contents Table of Contents
Previous Page  962 / 1096 Next Page
Information
Show Menu
Previous Page 962 / 1096 Next Page
Page Background

S946

ESTRO 36

_______________________________________________________________________________________________

Results

The optimisation process resulted in better image quality

in relation to the original presets and “standard images”.

Dose was reduced by a factor ranging from 2-4 times. For

a given mAs, superior image quality was seen for a higher

mA and lower ms, indicating that the detector response

was better for a higher dose rate. Saturation artefacts

(Fig.2) were visible for 64mA and 10ms when the images

included the intersection between the test object and air.

The worse UN was seen for LFOV. This was affected by

“cutting” from the reconstruction 40 pixel rows at the

edge of the panel. It was done because the bad pixel map

correction algorithm could not effectively correct the bad

pixels. Additionally, for 2D kV images, bad pixel artefacts

were visible using the TOR18FDG phantom. The kV

detector panel was replaced and the new one was

calibrated to get similar gains so the optimisation process

was still valid.

Conclusion

The performed optimisation process allowed us to manage

the image quality which met expected quality criteria

with significant reduction in dose.

EP-1724 Phantom image quality evaluation under 3

coil settings for abdominal MR-simulation at 1.5T

O.L. Wong

1

, J. Yuan

1

, S. Yu

1

, K. Cheung

1

1

Hong Kong Sanatoirum & Hospital, Medical Physics and

Research Department, Hong Kong, Hong Kong SAR China

Purpose or Objective

MR-simulation for abdominal radiotherapy often involves

the use of customized immobilization vacuum bags and

radiofrequency (RF) coil holders. Although several types

of RF coils are available for abdominal MR scans, the

influence of different RF coils and settings on image

quality has rarely been studied. In this study, we aimed to

quantitatively compare the quality of image acquired by

three different coil settings for abdominal MR-simulation

scan on a 1.5T MR-simulator.

Material and Methods

A

homogeneous

cylindrical

water

phantom

(diameter~21cm, length~35cm, volume~15L) was

positioned on a flat couch top

with a vacuum-bag. In

combination with a spine coil, three sets of scans, with 4

repeats each, were performed under the coil settings

(Fig1) with either a 18-channel body array (Body18x1), two

6-channel body arrays (Body6x2) or a single 6-channel

body array (Body6x1) on a dedicated 1.5T MR-simulator

(Aera, Siemens Healthineers, Erlangen, Germany). All

images were acquired using a 2D spin-echo T1-weighted

(TR/TE=500/20ms)

and

T2-weighted

(TR/TE1/TE2=2000/20/80ms) sequences (FOV=448mm,

matrix=448x448, slice thickness=5mm, geometric

distortion correction and prescan normalization=ON, 11

slices). For all scans, the coil-to-phantom distance

remained constant by fixing the coil holder height. SNR

was calculated based on AAPM Report 100 using the

central slice from each image set. For image uniformity

assessment, the percent of pixels with intensity within 10%

of the mean signal was calculated as uniformity index (UI).

A rank-sum test was performed to compare SNR and UI

differences between three coil settings.

Results

As illustrated in Fig2, the SNR of Body6x1 (T1:51.2±1.3,

T2:103.8±26.3) was significantly larger than that of

Body18x1 (T1:47.7±1.1, T2:81.9±6.7) for both T1 (P<0.01)

and T2 series (P<0.05). Compared to Body18x1, the SNR of

Body6x2 (T1:46.1±0.9, T2:96.7±10.5) was significantly

lower using T1 series (P<0.05) and larger using T2 series

(P<0.01). Significantly larger SNR of Body6x1 was also

noted comparing to Body6x2 using T1 series (P<0.01). For

image uniformity assessment, UI of Body6x1

(T1:92.8±0.6%, T2:89.0±0.3%) was significantly smaller

than Body18x1 (T1:96.4±0.6%, T2:82.5±0.2%) and Body6x2

(T1:96.0±0.2%, T2:82.6±0.2%) using T1 series (P<0.01),

and significantly larger than Body18x1 and Body6x2 using

T2 series (P<0.01). In terms of SNR and UI, Body6x1

outperformed other two settings for T2-weighted

abdominal MR-simulation. However, shorter coverage

along SI direction and smaller maximum acceleration

factor of Body6x1 might be a limitation for some

applications due to its smaller coil size and fewer array

elements.