![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0962.jpg)
S946
ESTRO 36
_______________________________________________________________________________________________
Results
The optimisation process resulted in better image quality
in relation to the original presets and “standard images”.
Dose was reduced by a factor ranging from 2-4 times. For
a given mAs, superior image quality was seen for a higher
mA and lower ms, indicating that the detector response
was better for a higher dose rate. Saturation artefacts
(Fig.2) were visible for 64mA and 10ms when the images
included the intersection between the test object and air.
The worse UN was seen for LFOV. This was affected by
“cutting” from the reconstruction 40 pixel rows at the
edge of the panel. It was done because the bad pixel map
correction algorithm could not effectively correct the bad
pixels. Additionally, for 2D kV images, bad pixel artefacts
were visible using the TOR18FDG phantom. The kV
detector panel was replaced and the new one was
calibrated to get similar gains so the optimisation process
was still valid.
Conclusion
The performed optimisation process allowed us to manage
the image quality which met expected quality criteria
with significant reduction in dose.
EP-1724 Phantom image quality evaluation under 3
coil settings for abdominal MR-simulation at 1.5T
O.L. Wong
1
, J. Yuan
1
, S. Yu
1
, K. Cheung
1
1
Hong Kong Sanatoirum & Hospital, Medical Physics and
Research Department, Hong Kong, Hong Kong SAR China
Purpose or Objective
MR-simulation for abdominal radiotherapy often involves
the use of customized immobilization vacuum bags and
radiofrequency (RF) coil holders. Although several types
of RF coils are available for abdominal MR scans, the
influence of different RF coils and settings on image
quality has rarely been studied. In this study, we aimed to
quantitatively compare the quality of image acquired by
three different coil settings for abdominal MR-simulation
scan on a 1.5T MR-simulator.
Material and Methods
A
homogeneous
cylindrical
water
phantom
(diameter~21cm, length~35cm, volume~15L) was
positioned on a flat couch top
with a vacuum-bag. In
combination with a spine coil, three sets of scans, with 4
repeats each, were performed under the coil settings
(Fig1) with either a 18-channel body array (Body18x1), two
6-channel body arrays (Body6x2) or a single 6-channel
body array (Body6x1) on a dedicated 1.5T MR-simulator
(Aera, Siemens Healthineers, Erlangen, Germany). All
images were acquired using a 2D spin-echo T1-weighted
(TR/TE=500/20ms)
and
T2-weighted
(TR/TE1/TE2=2000/20/80ms) sequences (FOV=448mm,
matrix=448x448, slice thickness=5mm, geometric
distortion correction and prescan normalization=ON, 11
slices). For all scans, the coil-to-phantom distance
remained constant by fixing the coil holder height. SNR
was calculated based on AAPM Report 100 using the
central slice from each image set. For image uniformity
assessment, the percent of pixels with intensity within 10%
of the mean signal was calculated as uniformity index (UI).
A rank-sum test was performed to compare SNR and UI
differences between three coil settings.
Results
As illustrated in Fig2, the SNR of Body6x1 (T1:51.2±1.3,
T2:103.8±26.3) was significantly larger than that of
Body18x1 (T1:47.7±1.1, T2:81.9±6.7) for both T1 (P<0.01)
and T2 series (P<0.05). Compared to Body18x1, the SNR of
Body6x2 (T1:46.1±0.9, T2:96.7±10.5) was significantly
lower using T1 series (P<0.05) and larger using T2 series
(P<0.01). Significantly larger SNR of Body6x1 was also
noted comparing to Body6x2 using T1 series (P<0.01). For
image uniformity assessment, UI of Body6x1
(T1:92.8±0.6%, T2:89.0±0.3%) was significantly smaller
than Body18x1 (T1:96.4±0.6%, T2:82.5±0.2%) and Body6x2
(T1:96.0±0.2%, T2:82.6±0.2%) using T1 series (P<0.01),
and significantly larger than Body18x1 and Body6x2 using
T2 series (P<0.01). In terms of SNR and UI, Body6x1
outperformed other two settings for T2-weighted
abdominal MR-simulation. However, shorter coverage
along SI direction and smaller maximum acceleration
factor of Body6x1 might be a limitation for some
applications due to its smaller coil size and fewer array
elements.