Table of Contents Table of Contents
Previous Page  972 / 1096 Next Page
Information
Show Menu
Previous Page 972 / 1096 Next Page
Page Background

S956

ESTRO 36

_______________________________________________________________________________________________

For each dynamic beam delivery the TrueBeam system

also produces a single log of the main system parameters

in a trajectory log text file. The TrueBeam Log Viewer

application can list these parameters versus time for user-

selected accelerators and data interval.

All lists can be sorted with respect to any column, non-

clinical events can be ignored, events not matching a user-

selected main fault code or fault description string can be

ignored, and all lists can be exported to Excel for further

analysis.

Results

Sorting the list of events with respect to the main fault

code provides a fast overview of the number faults for

each fault type and accelerator, facilitating an easy

prioritization of faults (see figure 1). In addition,

treatment plan details and mechanical axes for each event

(for example beam energy, the number of monitor units,

the patient ID, and the gantry angle) are valuable for the

identification of the root cause of the corresponding fault.

A more detailed insight into the root cause of a fault can

be gained by analyzing node records of the corresponding

event in Excel. An example of a node record for a gantry

fault is shown in figure 2. Also trending of the main system

parameters from the trajectory log text files with Excel is

a strong troubleshooting tool.

Conclusion

The TrueBeam Log Viewer application is an efficient tool

both for obtaining an overview of existing faults and for

identification of the root cause of faults. In our clinic the

application has provided a much more substantiated

prioritization of service tasks and faster identification of

the root cause of faults. Varian has no service tools with

this capability.

EP-1738 Performance of a new EPID panel and

opportunities for a fast MV-CBCT acquisition.

C. Kennedy

1

, C. Ling

2

, R. Scheuermann

1

, D. Mihailidis

1

, J.

Metz

1

1

University of Pennsylvania, Radiation Oncology,

Philadelphia, USA

2

Varian Medical Systems, Reserach and Development,

Palo Alto- CA, USA

Purpose or Objective

To study the performance, image quality and clinical

potential of a novel fast MV-CBCT panel mounted on a

straight-through linac with 6X-FFF (filter-free) beam.

Material and Methods

A prototype MV imager is mounted in-line with the linac

at 154 cm source-to-imager distance. MV-CBCT imaging is

performed with 28 cm fixed width (projected at isocenter-

100 cm) and variable field length (up to 26 cm max) FOV,

200º arc, 15 sec rotation time, at 5MU and 10MU dose

levels. Images of a Catphan® 604 phantom are analyzed in

terms of noise, uniformity, spatial and contrast resolution,

and contrast linearity. The results are compare with those

from Truebeam kV-CBCT imager, and potentially, a

Tomotherapy MV-CT imager, all benchmarked to

diagnostic CT scanner images. Imaging dose will also be

assessed for completeness of testing of the MV-CBCT

system.

Results

The uniformity and noise of the MV-CBCT was acceptable

but not as good relative to diagnostic CT and kV-CBCT,

with variation of 32HU from the center to the

periphery. Spatial resolution is shown in Fig. 1 with

3lp/mm for 5MU and 4lp/mm for 10MU dose levels, as

compared to 7lp/mm for both kV-CBCT and diagnostic

CT. The low contrast resolution of the MV-CBCT was >>

1%, compared to 1.0% and 0.5% for kV-CBCT and diagnostic

CT. Contrast linearity and sensitometry is shown in Fig. 2,

with the MV-CBCT (and 10MU dose level) being the

modality that deviates especially for higher density

objects like Acrylic, Delrin and Teflon (green data

points). Additional tests with varying field length (small,

medium and max length) for the MV-CBCT did not show

any correlation and/or improvement to the image quality

results. Imaging dose will be assessed and presented

based on absolute beam output calibration with a 6cc ion

chamber and OSLDs. The Tomotherapy data and analysis

is underway for comparisons.