![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0983.jpg)
S967
ESTRO 36
_______________________________________________________________________________________________
Conclusion
The ACDS is developing a comprehensive suite of audit
modalities aimed at ensuring patient safety across a range
of clinical practice. The new Level III (end-to-end) audit
joins the integrated range of multi-modality audits
provided by the ACDS including the Level Ib audit (on-site
linac output) covering reference beams, FFF and small
fields and the Level II audit (slab phantom combined with
array) audit covering conformal, IMRT, VMAT and FFF
treatments
plans.
EP-1756 Treatment planning and dosimetric validation
of bone oligomet SABR treatments on TomoTherapy
C. Thomas
1
, T. Burrows
1
, R. Lynn
1
, N. Milesi
1
, S. Petty
2
,
M. Stenson
1
, K. Blythe
1
, T. Greener
1
1
Guys and St Thomas NHS Foundation Trust, Medical
Physics, London, United Kingdom
2
Guys and St Thomas NHS Foundation Trust,
Radiotherapy Department, London, United Kingdom
Purpose or Objective
To establish whether the TomoTherapy helical delivery
system can accurately deliver high dose per fraction SABR
treatments to bone oligo-metastases within the NHS
England Commissioning through Evaluation SABR program.
Material and Methods
TomoTherapy Volo treatment planning system was used to
generate example SABR treatment doses of 10Gy and 15Gy
per fraction to a cylindrical PTV within a CT dataset of the
Delta-4 phantom. These treatments were delivered to the
Delta-4 phantom. Treatment plans for clinical oligo-
metastases in bone, with prescription doses of 27Gy/3#
and 30Gy/5# were generated and delivered to Delta-4
phantom, ionisation chamber and Gafchromic film.
Clinical treatment fractions were delivered in 2 half-
fractions in order to allow a mid-fraction imaging scan to
assess intra-fraction motion.
Results
Volo treatment planning system signalled when the
treatment planning objectives were not deliverable and
suggested modified treatment planning parameters. The
test cases measured on Delta-4 passed local gamma
analysis at 3%/3mm with >95% pass rate. Paddick and
CIRTOG conformity indices were improved with the use of
TomoTherapy compared with VMAT solution for the first
clinical case, and dose gradient between target and
critical structures was improved. For the first clinical case
measured on Delta-4, 100% of sampled detectors passed
within 3%/3mm gamma analysis and 98.5% passed within
2%/2mm.
Initial
transverse
EBT2
Gafchromic
measurement of the first clinical case showed satisfactory
qualitative agreement with treatment plan. Subsequent
EBT3 GAFchromic film measurements resulted in 97.5%
gamma passrate at 3%/3mm and mean dose deviation on
representative dose profiles of less than 2.2%. Average
intra-fraction motion between half fractions was 0.68mm
in X, 0.64mm in Y and 0.84mm in Z with standard deviation
of 0.62mm, 0.48mm and 0.79mm respectively.
Conclusion
GSTFT is the first centre with QA approval under the NHS
England CtE programme to treat bone oligometastatic
cases using the TomoTherapy treatment planning and
delivery system. Volo and Hi-Art systems are capable of
generating and accurately delivering homogeneous dose of
up to 15Gy per fraction in phantom studies. Clinically
approved treatment plans for bone oligomet cases
delivering up to 9Gy per fraction have been generated and
accurately delivered to diodes, ionisation chamber and
Gafchromic film. Intra-fraction motion was small and has
permitted the reduction of PTV margin from 4mm to 3mm.
EP-1757 QA of MLC Elekta Agility for Static fields
F. Tato de las Cuevas
1
1
Hosp. Univ. de Canarias, Medical Physics Dept., Santa
Cruz de Tenerife, Spain
Purpose or Objective
QA of MLC is one of the main points of any LINAC QA
program. Agility MLC (Elekta) have different properties
than most of the more common MLCs, like less interleaf
transmission. The objective is to perform the Agility MLC
QA in static mode using the electronic portal imaging
device (EPID) and make this process as fast and accurate
as possible.
Material and Methods
The LINAC is an Elekta Synergy with Agility MLC and 6 MV
photons. A software is developed in MATLAB with some
remarkable points:
1.
Elekta iCOMCAT software was employed to
generate and send the strip-test with multiple
segments as a unique treatment, as is much
faster than creating and irradiating a beam for
each segment. With the software of Elekta
iView is difficult to acquire a complete image
of each full segment as this is not fast enough,
so fluency corrections of these segments were
performed, in order to avoid erroneous pixel
values (PV) in the way: a) In a 23x23 open field
is acquired a horizontal profile and measure
the % PV (in the center position of each future
segment), this % is related to the PV of the
position of a reference segment. b) Measure
the mean PV in the center of each strip-test
segment, and obtain the % PV related to the
reference segment PV. c) Rescale the image of
each segment in order to obtain the % PV
(respect the reference segment). Finally make
the sum of all images.
2.
Segments of 2 x 20 cm (cross-plane x in-plane)
to form series of strip-test images with gaps
overlapping from 1.2 to 3 mm are acquired for
taking the MLC reference after calibration. The
strip-test need bigger gap spread than other
MLC in order to detect the gap position
correctly, because of the lower penumbra.
3.
To correct the collimator angle is used the
filtered back projection method, because is very
tricky to use the interleaf leakage, as this MLC
have much lower interleaf transmission than
other MLC, like Millenium (from Varian).
4.
To localize the radiation center (RC) of the EPID
is used a LINAC tray with centered radiopaque
mark. Four 20x20 fields are obtained with this
tray at 4 collimator angles. RC is determined for
gantry 0º detecting the mark position in each
image and obtaining the mean. A vector
displacement is created to obtain RC with one
image at 0º collimator. Tray images for various
gantry angles at 0º collimator are acquired, so
that with just one tray image is enough to detect
RC exactly. This method is faster than using
field edges, where at least 2 images at different
collimator angles must be acquired for each
gantry angle.
Measurements of leaf positions using light projection are
made. Also are obtained strip-test with films and analyzed
with
RIT
software.