Table of Contents Table of Contents
Previous Page  985 / 1096 Next Page
Information
Show Menu
Previous Page 985 / 1096 Next Page
Page Background

S969

ESTRO 36

_______________________________________________________________________________________________

Material and Methods

The leaf position is defined as the position of the 50% of

the dose profile. This measurement depends on relative

position of the beam source and of the leaves. In order to

validate the EPID measurements of the absolute leaf

positions, 10 dose profiles, at the center of different leaf

pairs of the same MLC field, were acquired with an Elekta

iViewGT EPID and with a diode positioned in a water

phantom. The comparison between the two detectors was

performed by Matlab. Garden Fence (GF) was chosen as

test of the leaf position accuracy and a preliminary study

on the gap width was conducted. Leaf position accuracy

was checked automatically with DC by acquiring GF at the

4 cardinal gantry angles and with all the beam energies (6,

10 and 15MV), while the reproducibility was tested with 5

GF repeated in one day and 6 repeated in a time interval

of 70 days.

Results

The difference between EPID and diode absolute

measurement of the leaf positions was less than 0.8mm

for all the analyzed leaves, resulting from the summation

of an error due to the isocenter identification (0,5mm)

plus the leaf positioning error (0.2mm). The gap width

study revealed that, because of the penumbra widening

observed in small fields, the leaf position could be

accurately measured as the 50% of the edge profile, only

if the gap width is equal or larger than 16mm with 6MV

beam. Therefore, GF with 20mm gap was chosen as leaf

position accuracy test for all the energies in order to

distinguish the effect of beam source from that of leaf

positioning. For the GF at different gantry angles the

difference between the measured and the prescribed

position was well within 1.0mm for all the leaves.

Moreover, reproducibility of each leaf position resulted to

differ from its average value less than 0.4mm.

Conclusion

This work permitted to assess the accuracy and the

repeatability of the Elekta Agility MLC leaf positioning by

the combined use of the Elekta IviewGT EPID and the

Dosimetry Check software through the acquisition and the

analysis of Garden Fence test. This system was validated

comparing the EPID with a diode in a water phantom and

assessing the minimum gap width necessary for an

accurate leaf position measurement at all energies which

is useful to distinguish issues related to beam symmetry

from those related to leaf positioning.

EP-1760 A simple method for estimating the

longitudinal isocentre shift due to gantry motion

R. Hudej

1

, D. Brojan

1

, S. Pulko

2

, P. Peterlin

1

1

Institute of Oncology Ljubljana, Department of

Radiophysics, Ljubljana, Slovenia

2

University Clinical Centre Maribor, Department of

Oncology- Radiotherapy Unit, Maribor, Slovenia

Purpose or Objective

The isocentre as a point of intersection of the three

rotational axes (gantry, collimator and treatment couch)

ideally remains fixed in space during the rotation of

gantry, collimator, or the treatment couch. Due to the

mechanical limitations, gantry sags slightly, and

consequently the radiation isocentre shifts slightly

towards the treatment couch when the gantry rotates

from the uppermost to the lowermost position. The

purpose of this study is to assess this shift.

Material and Methods

A strip of radiochromic film embedded in a suitable water-

equivalent phantom is irradiated with a cross-line half-slit

field from the top (0°). Then the gantry is rotated to the

lowermost position (180°) without moving the jaws and

the phantom is irradiated again. The film is scanned and

analysed with an image analysis script. The central lines

of both half-slit images are determined, then the

intersection angle between them is calculated, and finally

the distance between the intersections of extrapolated

lines with the 'sagittal” plane is calculated.

Results

This method was tested on 7 linacs of different makes and

models (Elekta Synergy Platform, Elekta Versa HD, Varian

Unique, Novalis Tx) in the authors' radiotherapy centres.

The average distance by which the isocentre moves

between both gantry positions was found to be 1.04 mm

(SD 0.30 mm), with the whole range covered by the [0.53,

1.48] interval. The two lowest values were achieved on

the two single energy Varian Unique linacs. It was found

out that the longitudinal isocentre shift is largely

independent of the gantry isocentre wobble determined

by the star-shot test. We also tested the alignment of the

collimator 0° setting with the gantry rotation plane. The

average deviation was found to be 0.16° (SD 0.10°), range

[0.04°, 0.31°].

Conclusion

The results appear consistent, but it would be helpful to

test the method on a wider pool of treatment machines

over a longer period of time. The longitudinal isocentre

shift during gantry rotation is a non-negligible parameter

which needs to be incorporated into the uncertainty

budget which is the basis for the CTV-PTV margin.

EP-1761 Workflow development for the clinical

implementation of an MR-guided linear accelerator

T. Stanescu

1

, A. Berlin

2

, L. Dawson

2

, J. Abed

2

, A.

Simeonov

2

, T. Craig

2

, D. Letourneau

2

, D. Jaffray

2

1

Stanescu Teodor, Radiation Physics- PMH, Toronto,

Canada

2

Princess Margaret Cancer Centre, RMP, Toronto, Canada

Purpose or Objective

Development of clinical workflows for the implementation

of a new external beam radiation therapy environment

which relies on hybrid MR-CBCT in-room imaging guidance.

Material and Methods

A standard radiation therapy 6X linac (TrueBeam, Varian

Medical System, Palo Alto, CA) was integrated with a 1.5

T diagnostic MR scanner (IMRIS, Minnetonka, MN). The MR

can move on rails and was tuned up to perform optimal

imaging inside the treatment room in the proximity of the

linac. The patient load is transferred directly between the

MR diagnostic table and the linac IGRT couch via a

hovercraft system (Zephyr XL, Diacor, Salt Lake City, UT).

No special MR safety requirements were employed

regarding the curation of the linac room – the linac/couch

can be freely operated mechanically when the MR magnet

is present - only typical MR room screening for

ferromagnetic content was implemented. Comprehensive

testing was completed to confirm negligible magnetic

field coupling between the MR and the TrueBeam system

(linac and patient table). Since the linac retains its default

features and an MR imager is available in the linac vault a

combined MR-kV approach can be employed for the

patient setup verification and treatment delivery. A new

software tool was developed in collaboration with Varian

to provide the computation and implementation of

treatment couch shifts based on soft-tissue information,

i.e. image matching between plan MR and guidance MR. In

this study, clinical workflows for liver and prostate were

developed and tested. Each site posed challenges from

patient image data planning and acquisition to RT planning

and in-room guidance. The approach was to integrate the

capabilities offered by the new technology in existing

processes.

Results

MR imaging protocols for planning and guidance were

established. The guidance scans were optimized to

minimize session time with negligible penalty on the

accuracy of the image matching process (planned vs. on