3M MDA 2

Listeria

species and

L. monocytogenes

Collaborative Study

OMA-2016-MONTH-XXX

5/4/16

The 3M MDA2 -

Listeria monocytogenes

method may generate

Listeria

monocytogenes

to levels sufficient to cause stillbirths and fatalities in

pregnant women and the immuno compromised, if exposed. The user must

train its personnel in current proper testing techniques: for example, Good

Laboratory Practices, ISO 17025, or ISO 7218.

Follow all instructions carefully. Failure to do so may lead to inaccurate

results. The 3M Molecular Detection Instrument is intended for use with

samples that have undergone heat treatment during the assay lysis step, which

is designed to destroy organisms present in the sample. Samples that have not

been properly heat treated during the assay lysis step may be considered a

potential biohazard and should NOT be inserted into the 3M MDS

(instrument).

Store the 3M MDA2 -

Listeria monocytogenes

at 2-8°C. Do not freeze. Keep

kit away from light during storage. After opening the kit, check that the foil

pouch is undamaged. If the pouch is damaged, do not use. After opening,

unused reagent tubes should always be stored in the resealable pouch with the

desiccant inside to maintain stability of the lyophilized reagents. Store

resealed pouches at 2-8°C for no longer than 1 month. Do not use 3M MDA

2 -

Listeria monocytogenes

past the expiration date. Expiration date and lot

number are noted on the outside label of the box.

After use, the enrichment medium and the 3M MDA 2 -

Listeria

monocytogenes

tubes can potentially contain pathogenic materials. When

testing is complete, follow current industry standards for the disposal of

contaminated waste. Consult local regulations for disposal.

5.2

Test Kit Reagents -

3M MDA 2 -

Listeria monocytogenes

- 96 tests

5.2.1

Lysis Solution (LS) tubes

. – 96 (12 strips of 8 tubes)

5.2.2

Listeria Reagent tubes

. – 96 (12 strips of 8 tubes)

5.2.3

Extra caps.

– 96 (12 strips of 8 caps)

5.2.4

Reagent Control (RC).

– 16 (2 pouches of 8)

5.2.5

Quick Start Guide

5.3

Additional Supplies and Reagents

5.3.1

3M Molecular Detection System (instrument), power supply, cord, USB

cable and software

5.3.2

3M Molecular Detection Speed Loader Tray

5.3.3

3M Molecular Detection Chill Block Tray and Chill block insert

5.3.4

3M Molecular Detection Heat Block Insert

5.3.5

3M Molecular Detection Cap/Decap Tool [Reagent]

5.3.6

3M Molecular Detection Cap/Decap Tool [Lysis]

5.3.7

Empty lysis tube rack

5.3.8

Empty reagent tube rack

AOAC Research Institute

Expert Review Panel Use Only

OMAMAN-29/OMAMAN-30 Combined Collaborative Study Protocol

OMA ERP - June 2016

ERP Use Only