portions, there were 0 presumptive positives out of 5 replicates. See Table 17 for robustness

1

results.

2

3

Independent Laboratory Observations

4

5

The results of this study demonstrate the ability of the MDA2

-

Listeria monocytogenes

to detect

6

the presence of

Listeria monocytogenes

in various food matrices and select environmental

7

surfaces after 24-28 hours of a primary enrichment. The MDA2

-

Listeria monocytogenes

offers

8

the benefits of extremely high sensitivity and high specificity for the detection of

Listeria

9

monocytogenes

while reducing the overall time to presumptive results. The MDA2

-

Listeria

10

monocytogenes

also reduces the total analyst time required to obtain final results by

11

implementing a single heat lysis step and cool step, followed up by a single transfer to molecular

12

reaction tubes.The small foot print of the Molecular Detection System requires only minimal lab

13

and bench space, making it easy to move the instrument or pair up additional MDS units to

14

perform multiple MDA runs concurrently. The updated software is user friendly with easy to

15

interpret results. The ability to examine Relative Light Unit (RLU) curves makes trouble

16

shooting more streamlined if any problems occur while testing is being conducted.

17

18

Discussion

19

20

There were no differences between the 24- and 26-hour enrichment time points for the

21

environmental surfaces tested, therefore a 24-hour minimum enrichment will be recommended

22

for these matrices.

23

24

Conclusion

25

26

The 3M MDA2 -

Listeria monocytogenes

AOAC PTM validation study included an

27

inclusivity/exclusivity, lot-to-lot/stability, robustness and matrix studies. The 3M MDA2 –

28

Listeriamonocytogenes

was compared to the USDA/FSIS-MLG 8.09, the FDA-BAM

29

Chapter 10, the AOAC 993.12 or the ISO 11290-1/A1 reference methods for the detection

30

of

Listeriamonocytogenes

. There were nosignificant differences between the 3M MDA2 -

31

Listeriamonocytogenes

method and the corresponding reference method for any of the

32

matrices evaluated. The 3M MDA2 -

Listeria monocytogenes

demonstrated reliability as a

33

rapid and sensitive method for the detection of

Listeria monocytogenes

for the following

34

matrices: beef hot dogs (25 g), deli turkey (125 g), queso fresco (25 g), vanilla ice cream (25

35

g), 4% milk fat cottage cheese (25 g), 3% chocolate whole milk (25 mL), whole melon,

36

romaine lettuce (25 g), bagged raw spinach (25 g), cold smoked salmon (25 g), raw chicken

37

leg pieces (25 g), sealed concrete (sponge enriched in 100mL), plastic (Enviro

sS

wab

38

enriched in 10 mL), stainless steel (sponge enriched in 225 mL).

39

40

References

41

42

1.

Food and Drug Administration Bacteriological Analytical Manual Chapter 10:

Detection and

43

Enumeration of Listeria monocytogenes in Foods

. February, 2013. (Accessed January 2015)

44

45

Formatted:

Font color: Auto