3

Place the 3M™ Molecular Detection Chill Block directly on the laboratory bench; (the 3M™

4

Molecular Detection Chill Block Tray is not used). Use the chill block at ambient laboratory

5

temperature (20-25°C).

6

7

P

REPARATION OF THE

3M™M

OLECULAR

D

ETECTION

H

EAT

B

LOCK

I

NSERT

8

Place the 3M™ Molecular Detection Heat Block Insert in a dry double block heater unit. Turn on

9

the dry block heater unit and set the temperature to allow the 3M Molecular Detection Heat

10

Block Insert to reach and maintain a temperature of 100 ±1°C.

11

12

NOTE: Depending on the heater unit, allow approximately 30 minutes for the 3M Molecular

13

Detection Heat Block Insert to reach temperature. Using an appropriate, calibrated thermometer

14

(e.g., a partial immersion thermometer or digital thermocouple thermometer, not a total

15

immersion thermometer) placed in the designated location, verify that the 3M Molecular

16

Detection Heat Block Insert is at 100 ±1°C.

17

18

P

REPARATION OF THE

3M™M

OLECULAR

D

ETECTION

I

NSTRUMENT

19

1.

Launch the 3M™ Molecular Detection Software and log in.

20

2.

Turn on the 3M Molecular Detection Instrument.

21

3.

Create or edit a run with data for each sample. Refer to the 3M Molecular Detection System

22

User Manual for details.

23

24

NOTE: The 3M Molecular Detection Instrument must reach and maintain temperature of 60°C

25

before inserting the 3M Molecular Detection Speed Loader Tray with reaction tubes. This

26

heating step takes approximately 20 min and is indicated by an ORANGE light on the

27

instrument’s status bar. When the instrument is ready to start a run, the status bar will turn

28

GREEN.

29

30

L

YSIS

31

1.

Allow the lysis solution (LS) tubes to warm up by setting the rack at room temperature (20-

32

25 °C) overnight (16-18 hours). Alternatives to equilibrate the LS tubes to room

33

AOAC Research Institute

Expert Review Panel Use Only

OMAMAN-29/OMAMAN-30 Combined Collaborative Study Protocol

OMA ERP - June 2016

ERP Use Only