Bacteria Plus kit. First, each lot of primer/probe was combined with each lot of Master Mix

and all combinations were tested with a single DNA Bacteria kit lot(see Table 6) on a single

Rotor-Gene Q instrument.

Table 6. Evaluation of

mericon

E.coliO157 Screen Plus and

mericon

E.coliSTEC O-typeLot-

to-lot Variability

Primer/Probes Lot 1

Lot 2

Lot 3

Lot 1

Lot 2

Lot 3

Lot 1

Lot 2

Lot 3

Master Mix

Lot 1

Lot 1

Lot 1

Lot 2

Lot 2

Lot 2

Lot 3

Lot 3

Lot 3

DNA Bacteria Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

Second, each of 3 lots of DNA Bacteria kits was tested with a single lot of

mericon

E.coliO157Screen Plus or

mericon

E.coliSTEC O-type PCR kits primer/probe and

master mix (see Table 7). At the same time, three Rotor-Gene Q instruments were tested

side-by-side with the same kit lots.

Table 7. Evaluation of DNA Bacteria Plus and QIAsymphony Lot-to-lot Variability and

Instrument Variability

Bacterial strains used for the

mericon

E. coliO157 Screen Plus lot to lot variation studies

wereE. coli strains O157:H7 (RM-HYG374), O157(RM-HYG381), O26:H (RM-HYG375), E.

coli O111:H- (RM-YG376)and

Citrobacter freundii

(ifp culture collection)

Bacterial strains used for the

mericon

E.coliSTEC O-type lot to lot variation studies were

E. coli strainsO157:H7 (RM-HYG374), O157(RM-HYG381),E.coli O26:H- (RM-HYG375),

E. coli O111:H- (RM-YG376),E.coli (O145 RM-HYG390)and

Citrobacter freundii

.

Stability testing: Shelf life

Product stability data was submitted with the QIAGEN

mericon

Salmonella study (July

2012). For the purposes of this study, shelf life stability studies are being submitted for the

E. coli

Screen Plusand

E. coli

STEC O-typeprimer probe sets, since these are the only

components of the workflow protocols that vary between the

mericon

protocols. The

Salmonella

summary data can be found in Appendix 2 of the current report. Bacterial

strains used were

Salmonellaenterica

subsp

enterica

serovar Abony

(NCTC

6017)

,

Salmonella enterica

subsp

enterica

serovar Enteriditis, (Sm24/Rif12/Ssq),

Citrobacter

freundii, (ifp culture collection).

mericon

E.coliO157 Screen Plus and

mericon

E.coliSTEC O-typepathogen kit stability

studies

Primer/Probes

Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

Master Mix

Lot 1

Lot 1

Lot 1

Lot 1

Lot 1

DNA Bacteria

Lot 1

Lot 1

Lot 1

Lot 2

Lot 3

Rotor-Gene Q

RGQ #1 RGQ #2 RGQ #3 RGQ #1 RGQ #1

OMAMAN-36 E/ AOAC PTM Report

ERP Use Only

January 2017

AOAC Research Institute

Expert Review Panel Use Only