1.

Transfer 0.5 grams L-Carnitine -d

3

to a 50 mL flask

2.

Dissolve and bring to volume with laboratory water

d.

Mixed Intermediate Internal Standard Solution (MIX-IS)

. Prepare fresh on the day of

analysis - must use same MIX-IS solution for both calibration working standards and

samples.

i.

Transfer 400 µL of each -choline-chloride-d

9

and 400 µL-carnitine-d

3

HCl stock

solution by a calibrated pipette into a 10 mL volumetric flask

ii.

Bring to volume with laboratory water and mix well.

e.

Intermediate Working Standards (IWS)

– Prepare fresh on the day of analysis

i.

Pipette 100 µL of the mixed intermediate Internal Standard solution (MIX-IS)

into three individual 50 mL volumetric flasks.

ii.

Using Table 1, add the required volumes of the stock choline and carnitine

solution by proper sizes calibrated pipets.

iii.

Bring to volume with laboratory water and mix well.

Table 1. Volumes of each Stock Solution for IWS

Stock Volume (µL)

MIX-IS (µL)

Final Volume (mL)

IWS -1

25

100

50

IWS -2

125

100

50

IWS -3

250

100

50

f.

Working Standard Solution (WS)

– Prepare fresh on the day of analysis. Use 90%

acetonitrile in final working solution.

i.

Pipette 200 µL of each Intermediate Working Standards (IWS-1, 2, and 3) into

three10 mL volumetric flasks

ii.

Add 800 µL laboratory water each volumetric flask

iii.

Bring to volume with acetonitrile and mix well

iv.

Transfer to autosampler vials for analysis.

Table 2. Concentration of Standards

Name

Choline

Choline-d

9

Carnitine Carnitine-d

3

Diluent

VF

µg/mL

mL

IWS-1

1.5

1.6

0.375

0.8

Water

50

IWS-2

7.5

1.6

1.875

0.8

Water

50

IWS-3

15

1.6

3.750

0.8

Water

50

ng/mL

90%

WS1

30

32

7.5

16

MeCN

10

WS2

150

32

37.5

16

MeCN

10

WS3

300

32

75

16

MeCN

10

Candidates for 2016 Method of the Year

327