16001n-MSS-MSS-SLV FOS01 AOAC997.08 Beneo-Orafti, 1 Febr 2016, page 11/18

3.2 Accuracy

The standard performance requirement 2014.002 stated that the recovery ranges between 90 and

110% over the range of the essay. This recovery was measured by spiking 9 different matrices at

different levels covering the proposed range from 0.05 to 5 g/100g reconstituted product.

The 3 raw materials available in the Test Materials Kit were used for this spiking.

Product

Spike Recovery

SPIFAN

CODE

spike

level Native

level *

*

Average

(%)

RSD

%

Spiked with

SRM 1849a

CLC10-b

5

<0,03

98,2

1,7 Orafti P95

infant powder

partially hydr Milk

based

1172572116

1

<0,03

99,8

0,4 Nutraflora P95

infant powder

partially hydr Soy

based

117257651Z 0,05

<0,03

101,4

1,7 Orafti HP inulin

Adult nutr powder

low fat

00394RF00 0,05

<0,03

107,1

7,2 Nutraflora P95

Infant elemental

powder

00403RF00

1

<0,03

101,5

3,5 Orafti P95

Infant formula powder

Milk based

D04HTCVV

5

<0,03

96,9

1,6 Orafti HP inulin

Infant formula powder

Soy based

E29JVLV

5

<0,03

98,4

5,5 Nutraflora P95

Infant formula RTF

Milk based

SPIFAN

Control milk

form

EV4H2R

0,05

<0,03

100,8

2,3

Orafti P95

Adult nutr RTF high

Protein

00730RF00

1

<0,03

98,8

1,2 Orafti HP inulin

Average

100.3

*Fructan results are expressed in g/100 g reconstituted final product.

The average recovery for the different levels of spiking ranges from 96.9 to 107.1%, for which

107.1% seems to be a higher value in the series of recovery results. It is clear that this value is for

spiking at the lowest level which generates the highest error but for this sample were also amounts

of fructans observed but below the LOQ.

The RSD ranges from 0.4 to 7.2%, for which the highest value is also the lowest spiking level.

The overall average recovery is very well, being 100.3% and meets the requirements of the SMPR

2014.002.

Detailed results can be consulted in annex 5.