The study experienced a number of issues that prohibit interpretation of results. Mold contamination of
NIST 2383 compromised the samples resulting in an average measurement ~1/2 of the certified value.
Also, only four of the eleven labs have replicate data, therefore repeatability cannot be reliably
calculated. For future consideration, the study needs to be repeated. A discussion about the forms
analyzed is missing in the manuscript. Intermediate Folic acid solution: gram is probably not the correct
unit. (2.5 g/100 ml). Also concentration is expressed in micro liter/mil. The report states that on
request of the General Referee a blank sample was included. However, AOAC's Collaborative Study
protocol describes the need of a blank sample. This should be corrected. The bias between the
certification value of IRMM CRM BCR 421 and the Biacore value needs to be explained.
This might be related to the fact that the certified value is based on total folate determined by
microbiological assay. However, when mentioned this should be properly described and discussed.
EXPERT REVIEW PANEL VOTE
AND
RECOMMENDATION
MOTION: Not to move forward to First Action Official Method
status.
Entered By:
Szpylka
Second By:
Konings
Vote:
7 Approved
0
Opposed
0
Abstain
M
OTION PASSED
ERP PROFILE SUMMARIES
191