3

3.2

Precision and Accuracy (Trueness)

Three samples were selected for precision studies:

-

Standard Reference Material (SRM) 1849a (National Institute of Standards and Technology

(NIST)).

-

Reference sample infant formula 04-1426 (International Bureau for Analytical Studies

(Bipea)).

-

Infant formula powder partially hydrolysed soy based (4, SPIFAN).

SRM 1849a and infant formula Bipea 04-1426 were selected for accuracy studies.

Samples were analysed in duplicate on six different days by two analysts.

Results are reported in Tables 4, 5 and 6.

Table 4: SRM 1849a: precision and accuracy (trueness).

SD(r)

RSD

r

SD(iR)

RSD

iR

RSD

r

Accuracy

His

315

319

5.62

2.69 0.8% 5.94 1.9% 101.2% yes

yes

Tau

36.6

35.5

0.76

0.42 1.2% 0.82 2.3% 97.0% yes

yes

Ser

720

708

8.91

5.16 0.7% 9.63 1.4% 98.4% yes

yes

Arg

400

404

14.31

3.2 0.8% 14.49 3.6% 100.9% yes

yes

Gly

241

249

1.47

1.5 0.6% 1.82 0.7% 103.3% yes

yes

Asp

1070

1089

16.17 8.24 0.8% 17.19 1.6% 101.7% yes

yes

Glu

2590

2675

80.1 21.71 0.8% 81.56 3.0% 103.3% yes

yes

Thr

640

644

9.07

5.08 0.8% 9.76 1.5% 100.6% yes

yes

Ala

455

448

5.74

3.48 0.8% 6.25 1.4% 98.5% yes

yes

Pro

1195

1181

11.69 7.24 0.6% 12.76 1.1% 98.8% yes

yes

Lys

1010

1054

10.74 8.73 0.8% 12.39 1.2% 104.3% yes

yes

Tyr

510

577

6.65

4.49 0.8% 7.37 1.3% 113.1% yes

no

Cys2

128.6

133

4.94

2.17 1.6% 5.17 3.9% 103.4% yes

yes

Met

n/a

467

5.53

3.98 0.9% 6.21 1.3% n/a

yes

n/a

Val

760

819

35.64

6.6 0.8% 35.95 4.4% 107.7% yes

yes

Ile

660

705

10.11 6.19 0.9% 11.01 1.6% 106.9% yes

yes

Leu

1261

1241

26.61 9.83 0.8% 27.5 2.2% 98.4% yes

yes

Phe

580

586

7

4.34 0.7% 7.64 1.3% 101.1% yes

yes

Meet SMPR spec.

Analyte

Ref. val.

[mg/100g]

Average

[mg/100g]

Precision

Recovery

SD(b)

AMINO-01

FOR ERP/WORKING GROUP USE ONLY