Background Image
Previous Page  122 / 162 Next Page
Basic version Information
Show Menu
Previous Page 122 / 162 Next Page
Page Background

EuroWire – March 2011

120

technical article

The GR-20 specification is as follows:

a) Loose tubes stored in a pedestal or

closure shall be capable of having a

minimum of 14ft of expressed buffer

tube stored in a pedestal or closure.

b) No greater than 0.15 dB average

attenuation increase at 1,550nm on all

fibres, during and after the last cycle.

Table 2

is an outline of these tests.

4 PE-90 as compared

to GR-20 and

FOTP-244 test results

4.1 20-ft PE-90 compared to 14-ft

FOTP-244 tube exposure

The current specifications outlined in

sections 2 and 3 have multiple differences

between them, as noted. This section

compares methods defined in the

PE-90 and GR-20 (referencing FOTP-244)

documents by comparing the resulting

attenuation loss using identical seg-

ments of cables from multiple cable

manufacturers.

Table 3

is a comparison of results generated

from testing identical cable samples to the

20-ft PE-90 method to the 14-ft FOTP-244

method. Four cable samples from two

different cable manufactures were tested

to both the five cycle PE 90 test method

and the two cycle FOTP-244 test method.

As can be seen in

Table 3

, the 20-ft PE-90

test consistently has higher maximum

attenuation changes at –40˚C than the

Telcordia GR-20/FOTP-244 14-ft test. The

magnitude of this attenuation increase is

seen to be in the range of two times higher

for the PE-90 method and specification.

As will be shown in the next section, the

main contributor to this attenuation loss

difference is the expressed tube length.

4.2 PE-90 as compared to FOTP-244

method

In order to better understand the

difference between the test methods of

PE-90 and FOTP-244, a 20-ft mid-span test

was completed on six different cables, to

each test method. The six cables that were

tested came from three different cable

manufacturers.

The 20-ft PE-90 test method and 20-ft

FOTP-244 method generally result in

similar fibre attenuation changes, as can

be seen in

Figures 1

and

2

.

Figure 1

shows the average attenuation

increases and

Figure 2

shows the maximum

attenuation increases. Out of the six cables

tested, four resulted in greater attenuation

loss with the FOTP test method over the

fifth cycle PE-90 method. While there

is variation in the results using the two

methods, when applying the RUS limit

of 0.1 dB maximum and 0.05 dB average,

the cable types generally result in either

passing or failing results regardless of the

method.

5 The effect of the

number of cycles

5.1 Cycles and maximum attenuation

peaks

The PE-90 test requires that the last of

at least five cycles be measured and

evaluated. As can be seen in the multiple

cycle PE-90 test data in

Figures 1

and

2

,

the greatest attenuation increases may

not necessarily occur during the fifth cycle

and may occur anywhere from cycle two to

cycle five.

In

Figures 1

and

2

, three out of the

six cables tested show the greatest

attenuation increases between the second

and fourth cycles, not the fifth cycle. By

only measuring the last cycle, attenuation

increases that may occur in the field during

the first one or two seasonal temperature

extremes may be missed during testing.

5.2 The effect of increasing the number

of cycles

In order to better understand the effect

of mid-span test temperature cycles on

attenuation losses, three cables were

Maximum attenuation increase (dB)

Cable 1 Cable 2 Cable 3 Cable 4 Cable 5 Cable 6

Cycle 1

Cycle 4

Cycle 2

Cycle 5

Cycle 3

20ft draft FOTP-244

Cycle Type 3

Average attenuation increase (dB)

Cable Type 1

Cable Type 2

Cycle Cycle Cycle Cycle Cycle Cycle Cycle Cycle Cycle Cycle

1 2 3 4 5 6 7 8 9 10

Figure 2

:

20-ft PE-90 and 20-ft draft FOTP-244 midspan test comparison – maximum attenuation increase

Figure 3

:

20-ft PE-90 mid-span test with five additional cycles – average attenuation increase

abl T 3