FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Cons/Weaknesses

Dionex ion chromatograph with suppressed conductivity measurement needs experienced analysts

Standard dilution given in the SMPR is not followed (homogeneity of the test sample given by using

2,5 g test material?)

Supporting Data

General Comment:

Method Optimization: no remark

Performance Characteristics:

NIST 1894a Infant formula was used as reference sample

Chloride:

Analytical Range:

4 - 2400 mg/100g, shown by 3 independent tests with independent

prepared standard solutions

SMPR: 5 – 500 mg/100g reconstituted final product

(corresponds to 20 – 2000 mg/100g powder)

Linearity:

R

2

= 0.9995

SMPR: not required

LOQ:

0.02 mg/100 g

SMPR: ≤ 5 mg/100g reconstituted product

Accuracy/Recovery:

94,7 % – 106 %

SMPR: 95 % – 105 %

Except of two data points the SMPR is met

NIST: 12 determinations: recovery: 103 – 110 % (7207 – 7677 mg/kg)

NIST 1849a reference fraction mass for chlorine: 7010 mg/kg +/- 170 mg/kg

Precision (RSD

r

):

NIST SRM 1849a: 1.90 %

1,22 % – 3,16 % for SPIFAN samples

SMPR: ≤ 2 %

Reproducibility (RSD

R

): has to be determined in a multi lab study

System suitability: checked by analyzing daily standards at lowest and midpoint of the analytical

l range. System recalibrated, when percent error of the check standard is > 5 %

SPIFAN samples: 13 fortified SPIFAN samples were analysed and validation data are provided

Recommendation:

The method should be formulated manufacturer neutral.

Method cannot be recommended at first action status as the precision and the accuracy/recovery data do

not meet the SMPR requirements as well as the test sample amount (should be 25 g in 200 ml). But the

method has potential to get first action status after some adjustments resp. improvements.

Chlor-03/Fluor-02 Review Forms