Table of Contents Table of Contents
Previous Page  456 / 1082 Next Page
Information
Show Menu
Previous Page 456 / 1082 Next Page
Page Background

S441

ESTRO 36 2017

_______________________________________________________________________________________________

7-field 6MV IMRT for a conventional Elekta Agility linac

(Elekta AB, Stockholm, Sweden), 6MV FFF single 360° arc

VMAT using Pinnacle 9.10 (Philips Radiation Oncology

Systems, Fitchburg, WI) for a non-MRL and CyberKnife

treatment using Multiplan (Accuray inc, Sunnyvale, CA).

Plans were acceptable if the 16 dose constraints of the

PACE trial (NCT01584258) were achieved, without a major

variation to the protocol.

Results

Clinically acceptable 7-field IMRT MRL plans (see Figure 1)

were achieved in all ten patients. Clinically acceptable

plans were also achieved for all ten patients using 9-field

IMRT, non-MRL 7-field IMRT, non-MRL VMAT and

CyberKnife treatment. Clinically acceptable 5-field IMRT

MRL plans were only possible in seven patients. Table 1

summarises the number of exceeded constraints, mean

rectal doses and mean bladder V37Gy for each plan type.

Given the small patient group, exploratory ANOVA

analyses were undertaken for the number of co nstraints

missed, the rectum D1cc and the two most challenging

constraints to achieve- rectum V36Gy and bladder V37Gy.

For the MRL, 5-field IMRT MRL plans performed

significantly worse in all these analyses compared to 7-

field IMRT. 7-field IMRT MRL plans had significantly lower

rectal doses compared to CyberKnife plans. No significant

differences were seen between 9-field IMRT MRL plans and

non-MRL VMAT plans compared to 7-field IMRT.

Conclusion

MRL IMRT plans for prostate SBRT achieved the PACE trial

constraints in all patients with 9-field appearing similar to

7-field IMRT. 5-field IMRT in this set-up appears inferior

for the MRL. All platforms could produce clinically

acceptable plans. Further work is needed for dosimetric

validation and feasibility of MRL delivery.

PO-0829 Robustness of IMRT and VMAT for

interfraction motion in locoregional breast irradiation

R. Canters

1

, M. Kunze-Busch

1

, P. Van Kollenburg

1

, M.

Kusters

1

, P. Poortmans

1

, R. Monshouwer

1

1

Radboud University Medical Center, Radiation oncology,

Nijmegen, The Netherlands

Purpose or Objective

Conventional techniques for locoregional breast

irradiation using field abutment are challenging, even

more in combination with breath-hold irradiation and with

hypofractionation, since over- or underdosage may occur

more consistently in the abutment region in these

circumstances. IMRT and VMAT techniques are likely to

result in more conformal and homogenous irradiation,

though robustness for anatomical and posture variations is

possibly an issue. Compared to conventional plans, the

beams are not fully opposing and fields cannot be opened

manually outside the outer contour of the breast and the

body. Therefore, in this study we evaluated the robustness

of both an IMRT and a VMAT technique for daily variations

in patient positioning in comparison to our conventional

technique.

Material and Methods

20 Patients treated with a dose of 16x2.66 Gy using a

conventional technique to the breast and axillary lymph

nodes levels I to IV (Figure 1a) were replanned with both

an IMRT and a VMAT technique using Pinnacle

autoplanning. The IMRT technique consisted of 6 beams

with 20

o

spacing, while the VMAT technique consisted of

opposing pairs of 24

o

arcs (Figure 1). The delivered dose

was calculated using the cone beam CT (CBCT) (Elekta XVI)

images for each fraction to quantify the influence of

patient positioning, both for an online and offline

correction protocol. Contours were transferred from

planning CT to CBCT by deformable image registration

using Mirada RTx. Density overrides were applied to

account for imperfections in Hounsfield unit values on the

CBCT. IMRT and VMAT techniques were compared to the

conventional technique for the V95%, conformity index

(CI), mean lung dose and mean heart dose. The CTV-PTV

margin used is 7mm. Since the setup error is accounted for

when evaluating dose on the CBCT, we used the CTV for

the evaluation.

Results

Evaluation of the treatment plans for 20 patients showed

that V95% coverage of IMRT and VMAT plans was

comparable to conventional plans (Table 1). Conformity

was significantly higher for IMRT and VMAT. Mean lung

dose was approximately 0.7 Gy lower on average, while

mean heart dose increased by approximately 0.7 Gy using

IMRT or VMAT. Robustness evaluation of the dose on daily

CBCT’s using an online positioning protocol showed that

V95% coverage remained stable for conventional, IMRT an

VMAT. Significant conformity improvement was obtained

using both IMRT and VMAT, and small differences in mean

heart dose (+0.7 G) and mean lung dose (-0.8 Gy) were

found. Evaluation of an offline positioning protocol

showed similar results.

Conclusion

Presented IMRT and VMAT techniques show a similar

robustness for interfraction motion in locoregional breast

irradiation compared to the conventional technique, while

conformity of the target volume is increased significantly.

An offline positioning protocol would be sufficient for

clinically acceptable set-up accuracy.