Table of Contents Table of Contents
Previous Page  503 / 1082 Next Page
Information
Show Menu
Previous Page 503 / 1082 Next Page
Page Background

S488

ESTRO 36 2017

_______________________________________________________________________________________________

radiotherapy.

PO-0894 Comparing the spatial integrity of 7T and 3T

MR images for image-guided radiotherapy of brain

tumors

J. Peerlings

1,2

, I. Compter

1

, F.M. Janssen

1

, C.J. Wiggins

3

,

F.M. Mottaghy

2,4

, P. Lambin

1

, A.L. Hoffmann

1,5,6,7

1

Maastricht University Medical Center+- GROW - School

for Oncology and Developmental Biology, Department of

Radiation Oncology - MAASTRO, Maastricht, The

Netherlands

2

Maastricht University Medical Center+, Department of

Radiology and Nuclear Medicine, Maastricht, The

Netherlands

3

Maastricht Brain Imaging Center - Scannexus, Maastricht

University, Maastricht, The Netherlands

4

University Hospital RWTH Aachen University,

Department of Nuclear Medicine, Aachen, Germany

5

University Hospital Carl Gustav Carus at the Technische

Universität Dresden, Department of Radiotherapy,

Dresden, Germany

6

OncoRay, National Center for Radiation Research in

Oncology, Dresden, Germany

7

Institute of Radiooncology, Helmholtz-Zentrum

Dresden-Rossendorf, Dresden, Germany

Purpose or Objective

In neuro-oncology, 3 Tesla (3T) MRI is the current clinical

standard for tumor localization, radiotherapy volume

delineation and stereotactic (radio)surgery, sometimes

complemented by amino acid PET imaging. With superior

SNR and image resolution, anatomical 7T MRI can visualize

micro-vascularization in glioblastomas potentially

allowing improved target volume delineation. However,

concerns about geometrical distortion (GD) with

increasing field strength (B

0

) are detrimental for

applications of 7T MR in image-guided interventions. For

high-precision treatment strategies, the spatial integrity

of anatomical images needs to be warranted within ±1mm.

The aim of the study was to evaluate B

0

- and sequence-

related GD in clinically relevant 7T sequences and

compare it to equivalent 3T sequences and CT images

Material and Methods

To quantify B

0

- and sequence-related GD in T1-GRE, T1-

TFE, T2-TSE, T2-TSE FLAIR on 7T and 3T sequences, a

dedicated anthropomorphic head-phantom (CIRS Model

603A) was used. The phantom is composed of bone-/soft-

tissue equivalent materials and contains a 3D grid (3mm

rods spaced 15mm apart). System-based distortion

correction methods were applied to restore the gradient

uniformity of 3T and 7T. For all CT and MR images, 436

points of interests (POIs) were defined by manual

reconstruction of the 3D grid points in the respective

images. GD was assessed in 3 ways. Firstly, global GD was

estimated by the mean absolute difference (MAD

global

)

between the measured and the true Euclidian distances of

all unique combinations of POIs, independent of location

within the phantom. Secondly, local GD was assessed by

MAD

local

between the measured and the true Euclidian

distances of all POIs relative to the magnetic field

isocenter. Thirdly, a distortion map was created by

evaluating 3D displacement vectors for each individual

grid point

Results

MAD

global

in 3T and 7T images ranged from 0.19−0.75mm

and 0.27−1.91mm, respectively, and was more

pronounced than in CT images. CT was not entirely free of

GD with MAD

global

ranging from 0.14−0.64mm. B

0

-related

GD was larger in 7T than in 3T MRI with MAD

local

ranging

from 0.21-1.81mm and 0.11-0.73mm, respectively

(p<0.05). MAD

local

increased with increasing distance from

the magnetic isocenter and largest GDs were noted at the

level of the skull (Fig. 1). MAD

local

was <1mm for all

sequences up to 68.7mm from the isocenter. Sequence-

related GD at 7T was prominent in T1-TFE and significantly

differed from other 7T sequences (p<0.001). Figure 2

shows an anisotropic distribution of GD in T1-TFE with

increasing GD along the frequency-encoding direction