Table of Contents Table of Contents
Previous Page  534 / 1082 Next Page
Information
Show Menu
Previous Page 534 / 1082 Next Page
Page Background

S519

ESTRO 36 2017

_______________________________________________________________________________________________

distinguish two regions between flap and film. In one

region, the film is at a distance of 5 mm from the

applicator, and in the other region at a distance of 7 and

9 mm (5mm of PMMA plus 2 or 4mm air gap

respectively).Two different treatment plans have been

designed, in the first one the source stops in the center of

the spheres and in the other one at the edge, to compare

the difference between dwell positions. The dwell times

are set to get the dose distribution as uniform as possible,

prescribing 6 Gy at a depth of 5 mm.

Results

Results obtained are shown in table 1. Underdosage is

observed, produced by air layers, ranging from 4.8% to

10.8% when dwell positions are at the center of the

spheres, and from 6.2% to 11.8% when dwell positions are

at the edge of the spheres, with 2 and 4 mm air gap

respectively.

Conclusion

In view of the results obtained, it can be concluded that

several layers of air between the applicator flap and the

skin can lead to considerable variation in dosimetry, which

may involve the loss of effectiveness of treatments with

this type of applicators. Thus, utmost care is required

during the placement of the flap to minimize the error due

to the air gap, therefore avoiding an underdosage in the

volume to be treated.

PO-0945 Pretreatment verification for brachytherapy

G. Fonseca

1

, M. Podesta

1

, M. Bellezzo

1

, B. Reniers

2

, F.

Verhaegen

1

1

Maastro Clinic, Physics, Maastricht, The Netherlands

2

University of Hasselt, NuTeC, Hasselt, Belgium

Purpose or Objective

Individual plan QA is not performed in brachytherapy

mostly due to the large uncertainty associated with dose

measurements. Traditional setups require precise and

accurate positioning, and therefore usually laborious

procedures to detect anything other than large

discrepancies with an unclear distinction between source

or detector mispositioning. This study evaluates the use of

an Electronic Portal Imaging (EPID) to verify the treatment

plan.

Material and Methods

The EPID panel response was characterized with an High

Dose Rate (HDR) Ir-192 source. A robotic arm was

employed for positioning within a water tank (Figure 1a)

assuring 0.2 mm accuracy during the calibration, which

covered a clinically relevant range for the distance

between the source and the panel (from 6 up to 25 cm).

Experiments were performed with an acquisition rate of

6.7 fps for a single catheter and for a gynecological

cylinder applicator (Figure 1b) with 5 catheters. Inter-

dwell distances of 2 and 5 mm were employed and the

experiments performed for source activities between 5

and 10 Ci. The EPID response is proportional to the source

activity so it is possible to obtain the activity by sending

the

source

to

pre-defined

dwell

position.

Results

3D Cartesian coordinates can be obtained with 0.2 mm

accuracy using a single EPID panel. The panel can clearly

identify dwell positions 2 mm apart even with the catheter

at 24 cm distance (Figure 1c) from the panel. Absolute

coordinates can be obtained by adding reference points

(representing the corners of the water phantom) in the

treatment plan that can be related with the position of

the water phantom over the panel during the experiments.

An

in-house

developed software compares all dwell

positions/times against the treatment plan. The software

can also monitor the sequence of the treatment

identifying the afterloader channel connected to each

catheter. Therefore, it is possible to detect catheter

misplacements, swapped transfer tube connections,

wrong dwell times and/or positions and also verify the

source activity.

Conclusion

This work describes an experimental system that can be

implemented in the clinic

providing experimental pre-

treatment verification that is not currently available. This

method provides several advantages when compared

against other dosimeters such as films or MOSFETs as it

combines a 2D dosimeter, which has an online response.

Our system can detect several problems that would be

unnoticed during the treatment if only traditional QA is

performed.

PO-0946 Entropic model for real-time dose

calculation: I-125 prostate brachytherapy application.

G. Birindelli

1

, J.L. Feugeas

1

, B. Dubroca

1

, J. Caron

1,2

, J.

Page

1

, T. Pichard

1

, V. Tikhonchuk

1

, P. Nicolaï

1

1

Centre Lasers Intenses et Applications, Interaction-

Fusion par Confinement Inertiel- Astrophysique,

Talence, France

2

Institut Bergonié Comprehensive Cancer Center,

Department of radiotherapy, Bordeaux, France

Purpose or Objective

This work proposes a completely new Grid Based

Boltzmann Solver (GBBS) conceived for the description of

the transport and energy deposition by energetic particles

for brachytherapy purposes. Its entropic closure and

mathematical formulation allow our code (M

1

) to calculate

the delivered dose with an accuracy comparable to the

Monte Carlo (MC) codes with a computational time that is

reduced to the order of few seconds without any special

processing power requirement.