Table of Contents Table of Contents
Previous Page  532 / 1082 Next Page
Information
Show Menu
Previous Page 532 / 1082 Next Page
Page Background

S517

ESTRO 36 2017

_______________________________________________________________________________________________

The number of needles varied from 0 to 8 (average 3.8

needles per application). The rescaled TRAK and mean

volume of the HR-CTV was 0.37 cGy m

2

(range: 0.23-0.50

cGy m

2

) and 26.8 cc (range: 8.0-59.1 cc), respectively. In

general, the TRAK value increased with volume. In figure

1a the TRAK values are plotted against the HR-CTV. The

relation between these parameters can be described by a

linear equation (see figure 1b). When setting an upper and

lower limit of two standard deviations a 95% confidence

interval can be derived and outliers can be identified. The

higher TRAK value of these outliers suggest the volume

that received the prescribed dose is much larger than the

HR-CTV. This was true for these plans: due to excessive

reduction of the HR-CTV, a higher dose in the IR-CTV was

desired and planned in the direction of the uterus top.

Conclusion

The HR-CTV can be used to predict the TRAK value.

Outliers may indicate abnormalities in treatment planning

and further inspection of their dose distributions is

required. In this study, the deviations in the dose

distributions of the outliers were accepted, since they

resulted in an improved individualized treatment plan.

Using this relationship, the quality assurance of the

treatment plan can be improved.

Poster: Brachytherapy: Physics

PO-0942 Real time in vivo dosimetry for cervix HDR

brachytherapy - feasibility study using a MOSFET

J. Mason

1

, P. Bownes

1

1

Leeds Cancer Centre, Medical Physics & Engineering,

Leeds, United Kingdom

Purpose or Objective

Implementation of in vivo dosimetry (IVD) in

brachytherapy is partly limited by lack of c ommercially

available devices that support IVD. In this study a

modified rectal retractor and MOSFET were used to

investigate the feasibility of real time IVD for cervix

brachytherapy with simulated treatment plans delivered

in a water phantom.

Material and Methods

A decommissioned rectal retractor was modified by

drilling a small hole to allow a microMOSFET to be

inserted. The MOSFET was commissioned measuring

energy dependence and angular dependence of response

for the range of source-MOSFET positions expected in

cervix brachytherapy treatments. Standard and conformal

cervix plans covering the range of applicator sizes and

geometries used in clinical treatments were delivered in a

water phantom. The MOSFET was monitored during

treatment delivery and measured doses compared to

treatment planning system (TPS) calculated doses for the

total plan and for ring and inter-uterine tube (IUT)

individually.

Results

Corrections were applied for energy dependence response

(6% variation between 1 and 8 cm source-MOSFET

positions) and angular dependence of response (up to 8%

under response for the largest polar angle of 170°). Total

plan measurements agreed with TPS calculated doses

within 3.1% - 7.8% for 30° and 60° applicators but

measured 16% -24% high for 45° applicators (k=2

uncertainty was estimated as 14% for total plan

measurements). Separate analysis of ring and IUT

measurements similarly showed good agreement for all

cases except the 45° IUT for which measurements were on

average 55.3% higher than expected. For the 45° IUT the

MOSFET position is directly in line with the source cable

and longitudinal source axis based on the source positions

assumed by the TPS (see figure). A combination of a small

rotation of the source relative to the IUT axis and

deviation of the actual source position from the centre of

the IUT could explain the measurement difference. To

verify this, treatments for the 45° applicator were re-

measured with the MOSFET taped to the outside of the

rectal retractor in a position that was not aligned to the

IUT and measured doses agreed within 8%.

Conclusion

In vivo dosimetry for cervix brachytherapy would be

feasible if commercial rectal retractors were designed to

allow a dosimeter to be inserted. However it is important

to avoid dosimeter positions aligned with the source

longitudinal axis as this is a region of high dose

uncertainty.

PO-0943 Evaluation of a recent in vivo dosimetry

methodology for HDR prostate BT using MOSFET

detectors

R. Fabregat Borrás

1

, S. Ruiz-Arrebola

1

, E. Rodriguez

Serafín

1

, M. Fernández Montes

1

, A. García Blanco

2

, J.