Table of Contents Table of Contents
Previous Page  781 / 1082 Next Page
Information
Show Menu
Previous Page 781 / 1082 Next Page
Page Background

S765

ESTRO 36 2017

_______________________________________________________________________________________________

delivered and accepted based on certificates provided by

the manufacturer and only minimum testing is performed.

At MedAustron, advanced acceptance testing procedures

of the QA equipment were additionally developed and

implemented by medical physicists. QA equipment passing

the advanced acceptance testing procedure was further

commissioned. This works consolidates preliminary

understanding for range and spot measurement equipment

(Grevillot et al, PTCOG2016), using additional data

obtained during beam delivery commissioning (e.g. long

term reproducibility of QA equipment). Additional

detectors were also commissioned, such as a 2D array of

ionization chambers and a diamond detector. The

commissioning methods developed allowed determining

the accuracy of the QA devices in clinical conditions and

better define the QA tolerances for periodic quality

assurance of the beam delivery system. The purpose of

this work is to guide medical physicists in the

implementation of dosimetry equipment and associated

phantoms as a pre-requisite to acceptance testing,

commissioning and further QA checks of the facility itself.

Material and Methods

A x-ray check was carried out for each ionization chamber

to verify the integrity of its construction. Positioning

accuracy of the water phantom scanning mechanism was

evaluated in 3D against laser tracker measurement. Range

measurement equipment was carefully calibrated by

measuring the entrance WET of the device. Long term

reproducibility of a multi-layer ionization chamber

detector used for daily QA of beam ranges was determined

and used to define morning QA tolerances of the beam

delivery system. Spot measurement equipment was

evaluated in terms of spot size, position and 2D

homogeneity against radiochromic films. Transverse

profiles acquired in water with diamond detector were

evaluated against pin-point detectors (Figure 1). An

ionization chamber-based 2D array was evaluated in terms

of effective depth of measurement, recombination, 2D

homogeneity and absolute dose against ROOS chamber.

Extensive commissioning procedures were developed

specifically for each piece of QA equipment, based on its

intended use.

Results

The advanced acceptance testing procedures developed

resulted in the identification of defects in several devices

from different manufacturers before clinical use. The

commissioning procedures allowed maximizing the

performance of the QA equipment and consequently the

quality of medical commissioning activities. Main overall

uncertainties

are

presented

in

Table

1.

Conclusion

The procedures implemented at MedAustron significantly

improved the knowledge and the performances of the

dosimetric equipment and therefore the quality of

medical commissioning activities at the facility. Our

experience shows that commissioning of QA equipment is

a necessary step towards high precision radiation therapy.

EP-1451 Validation of local tolerances for VMAT

patient specific QA using the IBA Compass system

E. Crees

1

, R. Hulley

1

, G. Kidane

1

, Y. Miao

1

1

Queen's Hospital, Department of Medical Physics,

Romford, United Kingdom

Purpose or Objective

The aim of this study was to review the locally set

tolerances for VMAT Head and Neck, Brain and Prostate

and Pelvic Nodes patient specific QA using the IBA

COMPASS MatriXX

evolution

measurements and computations

with the IBA COMPASS® system. Radiotherapy treatment

planning requires independent verification of both the

treatment planning system (TPS) dose calculation and the

patient dose delivery system. The verification of the dose

delivery system can be carried out independently to the

TPS dose calculation (in which case individual treatment

plans may be verified using a 2

nd

calculation alone) or can

be incorporated into individual patient specific

measurements. The pass/fail tolerances of patient

specific QA using the Compass system have been set

locally. The calculation method verification tolerance was

set for the comparison between the Compass dose

calculation and the treatment planning system

calculation. The measurement method verification

tolerance was set by comparing the MatriXX

evolution

measured dose to the treatment planning system

calculated dose.

Material and Methods

A retrospective study was performed on Head and Neck,

Brain and Prostate and Pelvic Nodes VMAT plans that were

produced on the Eclipse TPS using the AAA algorithm. Each

treatment plan was re-calculated using the Compass

software and the dose distribution of each plan was

measured using the Compass MatriXX

evolution

. Subsequently,

the Compass computed and measured dose distributions

were compared to the TPS calculated plan using gamma

index analysis. Bland Altman statistical analysis was

employed to compare gamma index results of the Compass

calculated and Compass measured dose distributions. The

analyses were performed based on the agreement

between the treatment planning system compared to the

measurement and Compass calculations. The tolerances

were set on absolute dose difference (2% for computation

and 3% for measurement on all points) and global gamma

index assessment (2%/2 mm criterion for 98% of points –

computation and 3%/3mm criterion for 95% of point -

measured).

Results

Across all treatment sites, the mean gamma index was

99.6% for the calculated dose and 98.3% for measured