Table of Contents Table of Contents
Previous Page  777 / 1082 Next Page
Information
Show Menu
Previous Page 777 / 1082 Next Page
Page Background

S761

ESTRO 36 2017

_______________________________________________________________________________________________

to optimize the dose sensitivity and the Signal-to-Noise

Ratio. In particular, geometrical distortion, image

homogeneity, artifacts, image texture, dose accuracy and

resolution, limit of detectability (LOD) and quantification

(LOQ), Fe

3+

spatial diffusion (Fricke-gels) and dose rate

dependence were evaluated. Finally, a pre-treatment

dosimetry of a SBRT plan was acquired and a relative

planar profiles comparison with a standard dosimeter

(Gafchromic EBT2) was performed. Ad hoc Matlab codes

were developed for images analysis.

Results

The chemical composition, MRI acquisition and

reconstruction parameters were optimized for each gel

system. No image correction maps were needed, since

geometrical distortion, artifacts and inhomogeneity were

always negligible, and no dependence on photon beam

dose rate was observed. 3D spatial resolution (voxel

dimension) was 1x1x3mm

3

. Dose accuracy was under 4% in

the range 18-25Gy, but worst for lower doses. Dose

resolution was about 1Gy, while LOD was less than 0.5Gy.

Differences between gel systems and Gafchromic profiles’

FWHMs were in the range 0,5mm – 5,5mm, mean dose

deviations in flat region were always around 2%, while

penumbra differences were about 2mm. Negligible

diffusion and time effects were observed up to 3 hours

from irradiation for all gel systems.

Conclusion

This study showed that both Fricke/Polymer- gel

dosimeter could be a suitable tool to perform pre-

treatment QA, with particular focus on SBRT and SRS

treatments, thanks to their optimal spatial resolution,

their practicability and their capability to perform 3D

dosimetry. Further studies are ongoing to standardize a

protocol to perform 3D pre-treatment dosimetry.

EP-1443 Measurement of 3D dose distributions from

an MR Linac with gel dosimetry

Y. Roed

1,2

, L. Pinsky

1

, G. Ibbott

2

1

University of Houston, Physics, Houston, USA

2

The University of Texas MD Anderson Cancer Center,

Radiation Physics, Houston, USA

Purpose or Objective

To demonstrate the potential value of polymer gels to

measure 3D dose distributions delivered with an MR-image

guided radiotherapy delivery machine.

Material and Methods

Polymer gels were obtained from MGS Research Inc

(Madison, CT) in custom-designed glass cylinders of 4 cm

height and 5 cm diameter. Irradiations were delivered

with a non-clinical MR-linac pilot system (MR-Linac, Elekta

AB, Stockholm) that combined a 1.5 T MR scanner with a

7 MV linac. Two dosimeters were positioned separately in

a phantom with their midplanes at isocenter distance. A

total of 750 MU (~5 Gy) was delivered with 3x3 cm² fields

at three gantry positions. The gantry was positioned at 0

,

90

, and 180

for the first irradiation and at 0

, 270

, and

180

for the second irradiation. All four cardinal angles

weren’t feasible due an asymmetric phantom design.

MR images across the entire volume of the dosimeter were

acquired with a 3T GE scanner using a 2D spin echo

sequence (TR = 1000 ms, TE = 10, 20, 60, 100 ms) 24h after

irradiation. Spin-spin relaxation rate (R2) maps were

generated. Both field size and penumbra widths were

calculated on the central slice. R2 maps were

concatenated into a 3D matrix.

The experiment was performed while the magnet of the

MR component (B-field) was turned off and will be

repeated once the B- field is turned back on.

Results

The small fields were captured and resolved within each

dosimeter. The field width measured along the central

cross-plane R2 profile from each dosimeter was 28 mm and

29 mm, respectively. The penumbra widths were 5 mm at

both field edges in each dosimeter. The 3D R2 matrix

visualized the irradiated volume of the dosimeter well.

In order to study the influence of the B-field on the dose

distribution in 3D, the results in the presence and absence

of the MR component (B-field) will be compared and

presented.

Conclusion

Polymer gels offer an excellent means to measure 3D

relative dose distributions delivered with an MR-Linac in a

clinically relevant fashion.

Previous experiments with polymer gels have already

shown that steep dose gradients could be measured when

irradiated with an MR-Linac. The current study encourages

further study of polymer gels for measuring 3D dose

distributions in the presence of B-fields.

EP-1444 Reliable error detection in radiochromic film

dosimetry with optimal density curves and corrections

H. Park

1

, Y. Bae

2

, J. Park

3

, M. Kim

1

, Y. Oh

1

, M. Chun

1

, O.

Noh

1

, O. Cho

1

, J. Lee

2

1

Ajou University Medical Center, Dept. of Radiation

Oncology, Suwon, Korea Republic of

2

Konkuk University Medical Center,

Dept.of

Convergent

Medical Physics and Dept. of Radiation Oncology, Seoul,

Korea Republic of

3

University of Florida, Dept. of Radiation Oncology,

Gainesville, USA

Purpose or Objective

To minimize variation of dosimetric errors caused by

correction methods and to suggest optimal conditions in

gafchromic film dosimetry using a flatbed scanner,

feasible scanning and post analysis procedures were

investigated with impacts on error detection in gamma

analysis.

Material and Methods

When a rectangular piece (5 × 4 cm) of EBT3 film was

placed at a 5 cm depth of the water-equivalent solid water

phantom, doses were delivered to film pieces from 0 cGy

to 20 Gy with every 50 cGy under 500 cGy and 100 cGy

over 500 cGy. To find an optimal sensitometric curve

having a large range of optical density (OD) and linearity

in doses of interest, a set of exposed films was scanned in

a flatbed scanner with different conditions by adjusting

brightness, contrast, and highlight from -50 to 50.

Sensitometric curves of a red and a green channel were

obtained with each scanning condition and used to

compare gamma distributions. In addition, to clarify the