Table of Contents Table of Contents
Previous Page  780 / 1082 Next Page
Information
Show Menu
Previous Page 780 / 1082 Next Page
Page Background

S764

ESTRO 36 2017

_______________________________________________________________________________________________

γmean of 0.29±0.13 with 100% of tests within alert

criteria, and a mean γ% equal to 96.9±5.2% with 96.0% of

tests within alert criteria. In contrast to our past

experience of patients with head-neck and pelvic

treatments, where the systematic use of IVD revealed

some discrepancies due to major anatomical variations or

random anatomical changes in terms of filling of

rectum/bladder, no relevant discrepancies were detected

in SBRT patient. The results are supplied in quasi real-

time, with IVD tests performed and displayed after only 1

minute from the end of arc delivery. Figure 1 shows the

SOFTDISO user interface.

Conclusion

The present EPID-based IVD algorithm provided a fast and

accurate procedure for SBRT-VMAT delivery verification in

clinical routine, with results obtained 1 minute after each

arc delivery. This strategy allows physics and medical staff

to promptly act in case of major deviations of dose

delivery.

EP-1449 The effect of a build-up screen on superficial

dose in total body irradiation

L.S. Fog

1

1

Rigshospitalet, The Clinic of Oncology, Copenhagen,

Denmark

Purpose or Objective

In total body irradiation (TBI), a build-up screen is

typically positioned between the linac and the patient to

reduce the build-up effect in the patient skin. With the

implementation of step and shoot TBI (SS TBI), the dose

conformity is considerably improved compared with TBI

delivered with open fields. Thus, the delivery of an

accurate skin dose becomes pertinent. We measured and

calculated skin dose for a range of TBI conditions.

Material and Methods

The dose was measured in a 20 cm thick solid water

phantom using a NACP parallel plate chamber, and a PTW

Unidos electrometer. The energy response from the

chamber contributed no more than 5% to the

measurement uncertainty (Phys Med Biol. 2001

Aug;46(8):2107-17).

The dose was measured at a depth of 1 mm in the

phantom; for a range of SSDs (340-440 cm); for 6 and

18MV; and for open jaw fields (used in conventional TBI

treatments) and MLC defined fields (used in SS IMRT); and

with and without a lucite build-up screen of 16 mm

thickness, placed 20 cm from the phantom. The MLC fields

were created with a 3 cm distance from the phantom edge

to the field edge when projected to the isocentre. The

chamber was calibrated by measurements under standard

reference conditions.

The doses were calculated using Eclipse™ (Varian Medical

Systems, Palo Alto), AAA algorithm, v.13.6, with a 1 mm

calculation grid.

The difference between measured and calculated doses

Δ

meas, calc

, between jaw and MLC fields Δ

jaw, MLC

, and with

and without build-up screenΔ

b,no b

were determined.

Results

For jaw fields, Δ

meas, calc

is reduced from 0-22% to 0-9%

when using a build-up screen (table 1).

For MLC fields, Δ

meas, calc

increases from 10-31% to 4-48%

when using a build-up screen.

With the exception of the scenario with jaws and a build-

up screen, Δ

meas, calc

increased with SSD.

The considerable Δ

meas, calc

values could arise from

inaccurate beam modelling in the build-up region; with

the modelling less accurate for MLC fields. For MLC

fields,Δ

meas, calc

increased with increasing SSD, perhaps due

to an underestimation of the scatter contribution from

MLC fields at extended SSD.

The jaw fields gave rise to a greater dose than the MLC

fields. The calculation underestimated Δ

jaw, MLC

with a

build-up screen, while it overestimated it without a build-

up screen.

Doses were greater with than without a build-up screen in

all the scenarios investigated. Δ

b,no b

was not considerably

different between the calculation and the measurement.

Δ

b,no

b

was greater for 18X than for 6X .

Conclusion

The presence of a build-up screen increases superficial

phantom dose. However, differences of up to 48% exist

between calculated and measured doses at the phantom

surface. These differences generally increase with SSD and

depend on beam energy and field type (jaw vs MLC) in a

complex way. The modelling of scatter from MLC fields at

large SSDs appears to be a particular challenge.

EP-1450 Implementation of dosimetry equipment and

phantoms in clinical practice of light ion beam

therapy.

L. Grevillot

1

, J. Osorio

1

, V. Letellier

1

, R. Dreindl

1

, A.

Elia

2

, H. Fuchs

3

, A. Carlino

4

, S. Vatnitsky

1

, H. Palmans

5

,

M. Stock

1

1

EBG MedAustron GmbH, Medical Physics, Wiener

Neustadt, Austria

2

EBG MedAustron GmbH / University of Lyon France,

Medical Physics, Wiener Neustadt, Austria

3

EBG MedAustron GmbH / Medical University of Vienna,

Medical Physics, Wiener Neustadt, Austria

4

EBG MedAustron GmbH / University of Palermo Italy,

Medical Physics, Wiener Neustadt, Austria

5

EBG MedAustron GmbH / National Physical Laboratory

UK, Medical Physics, Wiener Neustadt, Austria

Purpose or Objective

QA equipment (water phantoms, films, ionization

chambers, anthropomorphic phantoms, etc.) is generally