Table of Contents Table of Contents
Previous Page  789 / 1082 Next Page
Information
Show Menu
Previous Page 789 / 1082 Next Page
Page Background

S773

ESTRO 36 2017

_______________________________________________________________________________________________

shows comparable large possible beam width variations of

-17% to +25%, the maximum tolerable beam width

deviation in the entrance region amounts to -8% and +15%.

Conclusion

The QA plan in small depth shows rather small tolerable

deviations of the initial beam width. It is observed, that

this is mainly affected by the strong variation of the

particle numbers per scan spot in each energy slice,

optimized by the TPS to achieve lateral penumbras as

small as possible. Following this observation, we created

a plan with the same field size parameters consisting

uniform scanned energy slices. Applying the described

beam width variation, resulting tolerable beam width

deviations

are

-13%

and +12% .

For the QA plans in medium and large depth, beam width

tolerances in the SOBP become larger. One reason is that

the additional multiple coulomb scattering in water

dilutes the impact of deviations of the initial beam width.

Depending on the spot scan pattern, the entrance region

can be the region with the highest demands on beam width

accuracy for plans in large depths.

For TPS commissioning, the spot scan pattern should be

inspected with regard to beam spot width variations.

EP-1465 A beam matching procedure for Volumetric

Modulated Arc Therapy

L. Abdullah

1

, C. Constantinescu

1

, M.N. Hussein

1

1

King Faisal Specialist Hospital and Research center,

Biomedical Physics, Jeddah, Saudi Arabia

Purpose or Objective

To describe our experience of commissioning and quality

control tests, along with the adjustments performed for

beam matching in VMAT.

Material and Methods

A Synergy linac upgraded to Agility head was matched to

a Versa HD linac.

Dose calculation for commissioning tests was performed

on Monaco TPS version 5.0, with Monte Carlo algorithm,

inhomogeneity correction, calculation grid size of 1mm,

and statistical uncertainty of 0.5%/control point. The

compliance between dose calculation and measurement

was assessed for a gamma-index (GI) of 3% dose difference

and

3

mm

distance-to-agreement

(DTA).

The manufacturer specifies a routine for acceptance

testing of matched linear accelerators which concerns

only the beam quality and field size.

After appropriate MLC calibration, percentage depth dose

and beam quality index were evaluated for both linacs, in

similar setups. In-plane and cross-plane beam profiles

were acquired for field sizes of 10x10cm², 20x20cm

2

, and

depth of 10 cm. All dosimetric parameters appeared

identical for both linacs, within a tolerance of 1%. Beam

output calibration differed within 0.5%. The quality of

matching was found to be valid for 3D-CRT treatments but

not for VMAT. Specific test beams were further performed

to compare the two linacs for the accuracy of leave and

jaws movement, using the “picket fence” and “sliding-

window” methodology at different gantry and collimator

angles, using electronic portal imaging device (EPID)

dosimetry. The MLC minor leaf and jaw offsets of the

second linac was finely adjusted mechanically in order to

achieve acceptable accuracy of matching.

38 clinical VMAT plans for various sites and different

degrees of modulation were verified on both linacs. The

correlation between the GI passing rate and plan

modulation, assessed by the number of MU and number of

segments, was further investigated using a logistic

regression test.

Results

The beam quality index was 0.682 for linac 1 and 0.686 for

linac 2 after the vendor's matching procedure was

performed. GI analysis of the “picket fence” and “sliding

window” test beams indicated the need for mechanical

adjustment of MLC minor leaf and jaw offsets of linac 2.

After the beam matching accomplished, GI analysis of

VMAT clinical treatment plans indicated good agreement

between the two linacs. The average passing rate between

calculated and delivered dose distribution was 97.7±2.8%

(range: 87.5%-100%) for linac 1 and 93.4±4.7% (range: 83%-

99.%)

for

linac

2.

Weak correlations were found between the GI passing rate

and number of MU (R

2

=0.043, p=0.146) and segments

(R

2

=0.0273, p=0.240), indicating that the degree of

modulation is not to be considered in setting acceptance

criteria for GI analysis.

Conclusion

The manufacturer's defined criteria for beam matching are

not strict enough and should be evaluated before to

interchange patients between two matched linacs. We

described a method of evaluation and a procedure of beam

matching for VMAT

.

EP-1466 Implementation of a linac head-mounted

matrix detector to clinical use for dynamic technique

L. Szczurek

1

1

International Oncology Centre Affidea Poznan, Medical

Physics, Poznan, Poland

Purpose or Objective

Quality assurance of VMAT plans based on the results in 2D

phantom plane is not sufficient. Clinical interpretation of

2D measurements is difficult. The main purpose of this