Table of Contents Table of Contents
Previous Page  815 / 1082 Next Page
Information
Show Menu
Previous Page 815 / 1082 Next Page
Page Background

S799

ESTRO 36 2017

_______________________________________________________________________________________________

model. The highest couchtop attenuation (7.6%) was

measured at 135° gantry and 5×5 cm

2

field size. The

attenuation values of the H&N extension and breast boards

at 180° gantry angle were 6.9% and 6.7%, respectively. MC

results showed that the couchtop increased dose at

various depths of basal cell layer (0.1-0.4 mm) by 55.3%-

63.2%. The measured dose increase at 0.4 mm depth

ranged between 60.6% and 74.6% with field sizes 20×20

cm

2

to 5×5 cm

2

, the corresponding Co-60 unit increase for

a 10×10 cm

2

field being 18.1%. To directly compare two

prescribed treatment beams, when the PDDs were

normalized at 10 cm depth for a 10×10 cm

2

field, although

dose to subcutaneous tissues was always higher with the

Co-60 unit, it produced an at least 49.7% lower skin basal

layer dose.

Conclusion

The beam attenuation values should be applied in

treatment planning. The obtained skin dose results

support and explain the higher observed skin effects in

patients treated on the Compact unit compared to those

previously treated on the Co-60 unit with similar 180°

gantry angle beams. Modifying the treatment techniques

to reduce the fraction of the dose delivered through the

couchtop and/or the use of a ‘tennis racket’ type carbon

fiber couchtop should be considered.

EP-1509 Small fields defined by jaw or MLC: evaluation

of MU estimation by AAA and Acuros algorithms

F. Lobefalo

1

, A. Fogliata

1

, G. Reggiori

1

, A. Stravato

1

, S.

Tomatis

1

, M. Scorsetti

2

, L. Cozzi

2

1

Humanitas Research Hospital and Cancer Center,

Radiation Oncology, Milan-Rozzano, Italy

2

Humanitas Cancer Center and Humanitas University,

Radiation Oncology, Milan-Rozzano, Italy

Purpose or Objective

The small field output factor measurements are studied in

literature, covering the aspects of lack of charged particle

equilibrium, the partial occlusion of the finite source, and

the detector’s volume and response. However, the related

accuracy of the MU calculation from dose calculation

algorithms has not been investigated with similar

intensity. Aim of the present work is the evaluation of the

MU calculation accuracy for small fields generated by jaw

or MLC for two photon dose calculation algorithms in the

Eclipse system (Varian): AAA and Acuros. Simple static

beam geometries were chosen in order to better estimate

the accuracy with no additional biases. Flattening filter

free beams (6 and 10 MV) and and flattened 6MV were

evaluated.

Material and Methods

Single point output factor measurement were acquired

with a PTW microDiamond detector for 6MV, 6 and 10MV

unflattened beams generated by a Varian TrueBeamSTx

equipped with HD-MLC. Since the greatest

indetermination of the measurement accuracy resides in

the detector sensitivity correction factors for detector,

different corrections, field size dependent, were applied

according to different publications on the used detector.

Fields defined by jaw or MLC apertures were set; jaw-

defined: 0.6x0.6, 0.8x0.8, 1x1, 2x2, 3x3, 4x4, 5x5 and

10x10 cm

2

; MLC-defined: 0.5x0.5 cm

2

to the maximum

field defined by the jaw, with 0.5 cm stepping, and jaws

set to: 2x2, 3x3, 4x4, 5x5 and 10x10 cm

2

. MU calculation

was obtained with 1 mm grid in a virtual waterphantom

for the same fields, for AAA and Acuros algorithms

implemented in the Varian Eclipse treatment planning

system (version 13.6). Configuration parameters as the

effective spot size (ESS) and the dosimetric leaf gap (DLG)

were varied to find the best parameter setting.

Differences between calculated and measured doses were

analyzed.

Results

Agreement better than 0.5% was found for field sizes equal

to or larger than 2x2 cm

2

. In the following the results are

given for the two extreme detector sensitivity correction

factors, with the second value in brackets.

A dose

overestimation was present for smaller jaw-defined fields,

with the best agreement, over all the energies, of 1.6

(0.5)% and 4.6 (3.5)% for a 1x1 cm

2

field calculated by AAA

and Acuros, respectively, for a configuration with EES=1

mm for X, Y directions for AAA, and EES=1.5, 0 mm for X,

Y direction for Acuros. Conversely, a calculated dose

underestimation was found for small MLC-defined fields,

with the best agreement averaged over all the energies,

of -3.9 (-4.9)% and 0.2 (-0.8)% for a 1x1 cm

2

field

calculated by AAA and Acuros, respectively, for a

configuration with EES=0 mm for both directions, both

algorithms.

Conclusion

For optimal setting applied in the algorithm configuration

phase, the agreement of Acuros calculations with

measurements could achieve the 3 (6)% for MLC-defined

fields as small as 0.5x0.5cm

2

. Similar agreement was found

for AAA for fields as small as 1x1 cm

2

.

EP-1510 Dosimetric characterisation of stereotactic

cones by means of MC simulations

A. Nevelsky

1

, E. Borzov

1

, S. Daniel

1

, R. Bar-Deroma

1

1

Rambam Medical Center, Oncology, Haifa, Israel

Purpose or Objective

The objective of this work was to employ an MC model of

6MV FFF beam from the ELEKTA VersaHD linac to perform

dosimetric investigation of the new ELEKTA stereotactic

cones.

Material and Methods

The BEAMnrc code was used to create detailed model of

the linac head and stereotactic cones for the 6MV FFF

beam based on the manufacturer data supplied by Elekta.

MC simulation with the BEAMnrc code generated the

phase-space file which was used in the DOSXYZnrc code to

calculate PDDs, lateral profiles and output factors in a

water phantom for stereotactic cones with 5, 7.5, 10, 12.5

and 15 mm nominal diameter. Results from the

simulations were compared against measurements

performed in water phantom with PTW PinPoint ion

chamber and Scanditronix stereotactic diode. Actual cone

diameter was found by the best match between the

calculated and measured lateral profiles. Sensitivity of

output factor to cone diameter variations was

investigated. For this purpose, nominal cone diameter was

changed by +/- 0.3 mm (which is twice the manufacturer

stated uncertainty of 0.15mm).

Results

Lateral profiles agreed within 2%/0.5mm for all cone sizes.

Actual cone diameters were found to be 5.30, 7.70, 10.15,

12.65 and 15.15 mm. For the actual cone diameter, output

factors agreed within 2% for all cones except for cone 5

mm where the difference was 4%. Cone diameter

uncertainty of 0.3 mm lead to up to 11% variation in the

output factor compared to output factor value calculated

for the nominal diameter.

Conclusion

The MC model of the VersaHD linac was employed for

investigation and characterization of stereotactic cones.

Measured data were verified by the MC calculations.

Differences between nominal and actual cone diameter

were observed. Given the level of manufacturing accuracy

and sensitivity of dosimetric parameters to the cone

diameter variation, accurate commissioning of

stereotactic cones must be performed and comparison

with the data from other centers may be misleading.

EP-1511 Radiation Dose from Megavoltage Cone Beam

Computed Tomography for IGRT

E. Kara

1

, B. Dirican

2

, A. Yazici

1

, A. HICSONMEZ

1

1

Onko Ankara oncology center, Oncology Department,

Ankara, Turkey