S882
ESTRO 36 2017
_______________________________________________________________________________________________
procedure can be performed if an indisputable evidence
of improvement may be proved and the feasibility in a
clinical workflow is guarantee. In this work IVD performed
with electronic portal imaging device (EPID) was used to
evaluate VMAT and IMRT breast treatment performed with
a new set up and immobilization procedure.
Material and
Methods
IVD with EPID was performed over 32 patients that
underwent an IMRT or VMAT breast plus supraclavicular
treatment. Half of the patients followed the standard set
up procedure (SP) of the department, consisting of a
thermoplastic mask covering the district to be treated,
patient marks over the mask, bolus applied over the mask;
the others followed a new procedure (NP) and were
immobilized with a breast board and a knee support,
patient marks over the skin and bolus applied over the
skin. The accuracy of the treatment was evaluated with a
commercial software (SOFDISO, Best Medical Italy) that
provided two indexes: the ratio R between the
reconstructed (D
iso
) and planned (D
tps
) isocenter dose
(R=D
iso
/D
tps
) which can represent the accuracy of the dose
delivered, and a Pγ% obtained performing a gamma
analysis between the first EPID image and the next ones
acquired immobilized. Three consecutives tests were
scheduled during the first week of treatment and
successively two IVD test per week. The MLC log files of
the treatments delivered where analysed with a
commercial software and compared with the planned
treatment in order to discriminate the deviation coming
from the patient (anatomy and set up), from the deviation
coming from the linac.
Results
Only the IVD test coming from a delivery with the
machines log file in tolerance were considered. The
results of 545 IVD tests obtained over 32 patients were
reported in Table 1. Every treatment performed with IMRT
and VMAT resulted in 100% of the patients with R and Pγ
indexes in tolerance as for SP as for NP. The percentage
of Pγ index in tolerance as for VMAT as for IMRT increased
with NP. A 10% of, off tolerance tests persisted. The IVD
tests off tolerance were reported in the acceptable
threshold before the next fraction.
Discussion:
the new
immobilization procedure enabled a direct localization of
the patient skin and of the bolus positioned over it. The
use of the mould mask, positioned over the patient’s lead
to a non-direct evaluation of the patient rotation and
accommodation inside it. The beam can lack of
reproducibility if considering its path: air gap between the
bolus, the mask and the patient skin not considered in the
treatment planning. This aspect is moreover important for
IMRT treatment where for some beam entry this situation
can be more evident.
Conclusion
IVD in is a powerful tool that can be helpful in the
validation of new set up and immobilization procedures.
EP-1648 Thermoplastic mask dependency with
interfractional uncertainties for head and neck VMAT
treatments
E.M. Ambroa Rey
1
, R. Gómez Pardos
1
, D. Navarro
Giménez
1
, A. Ramirez Muñoz
1
, M. Colomer Truyols
1
1
Consorci Sanitari de Terrassa, Medical Physics Unit-
Radiation Oncology Department, Terrassa, Spain
Purpose or Objective
Volumetric-modulated arc therapy (VMAT) techniques
have the ability to deliver a highly conformal dose
distribution to the target and high dose gradient at the
interface between the tumor and the normal tissues,
decreasing the irradiated volume and sparing OARs.
However inaccurate alignment of the radiation beam with
the patient can lead to critical organs to receive an
unwanted high dose or the tumor to receive a reduced
dose producing a loss in tumor control. Radiation therapy
for head-and-neck (H&N) cancer requires a reliable
immobilization
for
an
accurate
treatment.
The purpose of this study is to establish the interfractional
setup error for VMAT H&N patients, using a kilovoltage
cone beam CT (CBCT) and a robotic treatment couch
(HexaPOD) for accurate patient positioning in six degrees
of freedom and analyze the differences between two
types of thermoplastic masks (Head mask (H) and Head
and Shoulder (HS) mask).
Material and Methods
Dataset for a total of 945 CBCT scans were obtained from
30 patients (13 with H mask and 17 with HS mask). For
each fraction, patients were placed on the HexaPOD
robotic couch and a CBCT was performed. For
interfraction accuracy evaluation, the daily variations of
the three principal axes (X, Y and Z) and three rotational
movements (pitch, roll, and yaw) were extracted. Also,
the type of thermoplastic mask was recorded. The
following parameters were calculated: the mean of the
setup corrections (M), the standard deviation (random
error, σi), the group random error (σ) defined as the mean
of all the individual patient random error σi and the
systematic group error (∑) defined as the standard
deviation of all the means measured for each patient.
Results
Results are shown in Table I. The overall mean
displacements are larger for the HS mask patients than for
the H mask. Also a trend toward a positive antero-
posterior direction (1.19 mm) was observed in
translational displacements for the HS mask patients. This
could be explained taking into account the differences
between the simulation and treatment couches as well as
the accuracy of the room lasers. The remaining error
components
did
not
show
any
trend.
For the rotational directions, the bigger error was in the
pitch direction for both the H mask (0.76) and HS mask
patients (0.85).
Regarding the systematic group error, we have found a
larger error for the HS mask patients (2.34) again toward
a positive antero-posterior direction. The random group
error shows the same behavior as well.
Conclusion