Background Image
Table of Contents Table of Contents
Previous Page  90 / 142 Next Page
Information
Show Menu
Previous Page 90 / 142 Next Page
Page Background

3/17/2015

12

LC CONDITIONS

Waters Acquity ACQUITY UPLC

®

BEH C18 1.8 µm, 2.1 x 100 mm

column

Mobile phase A: 10 mM ammonium acetate in water

-

Mobile phase-B: 10 mM ammonium acetate in 95/5 (vol.)

acetonitrile/water

Gradient conditions

Chromatography 

Flow Rate 

300 µL/min 

Column Temperature 

40°C 

Injection Volume 

10 μL  

S l T

A bi

Proprietaryand confidential — do not distribute

© 2015 Abbott Laboratories

March17,2015

MFA‐AOAC, 2015 

23

amp e emperature 

m ent

Gradient Program 

Time 

% A 

% B 

Curve 

0.00 

90 

10 

1.00 

90 

10 

7.00 

45 

55 

9.00 

100 

9.01 

90 

10 

12.00 

90 

10 

MS CONDITIONS

MS Tune  

Ionization Mode 

ESI‐ 

Cone Gas 

250 L/Hr 

Capillary 

0.5 kV 

Nebulizer 

7.0 bar 

Source Offset 

20.0 V 

Collision Gas Flow

0.15 mL/min

Source Temperature  150°C 

Quad 1 Resolution 

Unit Mass (0.75 Da FWHM) 

Gas Temperature 

350°C 

Quad 2 Resolution 

Unit Mass (0.75 Da FWHM) 

Desolvation Gas Flow  900 L/Hr 

MS Calibration Range 

50‐2000 amu 

MS/MS Transitions 

Compound 

Parent Mass (m/z) 

Daughter Mass (m/z) 

Dwell (s) 

Cone (V) 

Collision Energy (V) 

MFA‐2NPH 

212 

182 

0.15 

20 

15 

MFA‐2NPH_IS  

216 

186 

0.15 

20 

15 

Proprietaryand confidential — do not distribute

© 2015 Abbott Laboratories

March17,2015

MFA‐AOAC, 2015 

|

24