![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0333.png)
A Practical Example: SRT Case
Intracranial lesion: 3 x 8 Gy @ 80%
N=3,
β
=0.84
SD of the penumbra is 3.2 mm
σ
pen
=3.2 mm
E2E test device error (
Σ
) = 0.4 mm
Σ
1
=0.4 mm
Localization (delineation) error = 1.0 mm (1 SD)
Σ
2
=1.0 mm
Systematic error = 0.5 mm (1 SD)
Σ
eff
=0.58 mm
Random error = 0.5 mm (1 SD)
σ
eff
=0.41 mm
Intra-fraction error = 0.5 mm ( 1 SD)
σ
eff
=0.20 mm