![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0336.png)
A Practical Example: SBRT Lung Case
T1 primary lung lesion: 3 x 18 Gy @ 80%
N = 3,
β
= 0.84
Alignment on time-averaged tumor position by CBCT
SD of the penumbra is 6.4 mm
σ
pen
= 6.4 mm
E2E test device error (
Σ
) = 0.4 mm
Σ
1
= 0.4 mm
Localization (delineation) error = 2.0 mm (1 SD)
Σ
2
= 2.0 mm
Systematic error = 1.0 mm (1 SD)
Σ
eff
= 1.0 mm
Random error = 1.0 mm (1 SD)
σ
eff
= 1.0 mm
Intra-fraction amplitude = 1 – 25 mm
σ
r
= 0.4 – 9.0 mm