Previous Page  173 / 204 Next Page
Information
Show Menu
Previous Page 173 / 204 Next Page
Page Background

Analog Panel Meter, Rectangular

Part Number

Range

Type External

Resistance

Accuracy

Dimensions

M10-K

0 to 2500°F

or

-20 to 1370°C

K

10 Ohm

± 1%

or

25°F

Diameter -

4.50” or 114.3 mm

Depth -

1.50” or 38.1 mm

M10-J

0 to 1000°F

or

-20 to 540°C

J

10 Ohm

± 1%

or

10°F

M11-J

0 to 500°F

or

-20 to 260°C

J

10 Ohm

± 1%

or

5°F

2

M10 & M11

Analog Panel Meter, Round Portable Type -

Part Number

Range

Type

External

Resistance

Accuracy

Dimensions

M1

0 to 2500°F

or

0 to 1370°C

K

1 Ohm

± 2%

or

50°F

Diameter -

4.50” or 114.3 mm

Depth -

1.50” or 38.1 mm

1

M1

Richards Analog Panel Meters are available for both Type K and Type J thermocouples. It is important to note

that the

M10-K

,

M10-J

, and the

M11-J

, have an external resistance requirement of 10 Ohms. In order to main-

tain the accuracy of the readings, the total resistance of the thermocouple and the extension wires combined must

be equal to 10 Ohms. The

M1

unit has an 1 Ohm external resistance requirement. This requires that the total re-

sistance of the thermocouple and the extension wires combined must be equal to 1 Ohm.

The

M1

meter is normally used in our portable molten metal handle assembly, but they may be used in situations

where long extension wires are not required. The M10 and M11 type instruments are usually used in situations

where longer extension wires are needed.

! Note: Resistance values for thermocouple and extension wires can be found in the engineering section of this catalog.

Switches, Gold Fuses, Instruments - Analog, Round Portable or Square Panel

Manufacturing, Engineering and Sales Since 1938

Phone: 617.527.4385 Fax: 617.964.3746 Email:

sales@asrichards.com

SW-GF

M1-M10

Page 173

Visit our website at

www.asrichards.com