Background Image
Table of Contents Table of Contents
Previous Page  107 / 116 Next Page
Information
Show Menu
Previous Page 107 / 116 Next Page
Page Background

PDE5-005

Question

Response

General

AOAC Method Number:

PDE5-005

Method Title:

SLV Study of a method for Screening

Method Author(s):

Vaclavik, Schmitz, Halbardier, Mastovska

Reviewer Name:

Tom Phillips

Summary of Method

PDE5 inhibitors are extracted from the dietary ingr

edient or supplement samples using a 50:50 (v/v

) mixture of acetonitrile and water and centrifuged

. Supernatant is diluted, filtered and analyzed by LC-

HRMS. Data are collected in MS acquisition mode that

combines full-scan MS experiment with ion

fragmentation and data-dependent MS/MS product ion

scan experiment. This approach enables collection of MS

and MS/MS data for both targeted and non-targeted

PDE5 inhibitors in a single chromatographic run.

Software-facilitated identification of targeted analytes is

performed based on retention time, accurate mass and

isotopic pattern of pseudomolecular ion, and accura

te masses of fragment ions using in-house developed

compound database. Detection and identification of

other PDE5 inhibitors and novel analogues are performed

by retrospective evaluation of MS and MS/MS

experimental data.

Method Scope / Applicability:

Applicable to screening and identification of acet

aminotadalafil, acetildenafil, avanafil, homo sildenafil,

hydroxyacetildenafil, hydroxyhomos ildenafil,

hydroxythiohomosildenafil, lodenafil carbonate,

mirodenafil, propoxyphenyl homohydroxysildenafil,

sildenafil, tadalafil, thiohomosildenafil, udenafil,

vardenafil and other known and novel analogues of the

above PDE5 inhibitors.

General Comments About the

Method:

The method is miniaturized which allows high throughput

for samples. Sample preparation is easy and simple. The

linearity for the three compounds studied is excellent,