Background Image
Table of Contents Table of Contents
Previous Page  104 / 116 Next Page
Information
Show Menu
Previous Page 104 / 116 Next Page
Page Background

PDE5-005

of drugs to obtain structurally significant fragment ions

Performance Characteristics

Analytical Range:

0, 100, 1000 mg/kg

LOQ:

N/A

Accuracy/Recovery

Recoveries of a 50 ppm spike ranged from 69% to 90%.

Precision (RSDr)

Repeatability values obtained were 0.4-1.8%.

Reproducibility (RSDR)

N/A

System Suitability

Not addressed in method.

Safety Review

Is the method safe to use as written?

Yes

Are there any safety concerns with

any component, instrument, reagent

or instruction of the method? If yes,

please clearly state the concerns.

No

Does the method include all

necessary safety precautions and/or

warning statements? Yes or No. If

No, please suggest potential

precautions or warnings that can be

added to the method for users or

recommend safer option(s) if

warranted.

Yes

Recommendations

Final Recommendation:

Method is recommended for approval for screening,

identification and quantitation with reservation due to

expensive, highly technical instrumentation employed.