CHON-002
LOQ:
9 mg/L
Accuracy/Recovery
99.7 +/- 2.6%
Precision (RSDr)
< 2%
Reproducibility (RSDR)
< 2% reported
System Suitability
Not explicitly addressed in paper.
Safety Review
Is the method safe to use as
written?
Yes
Are there any safety concerns
with any component, instrument,
reagent or instruction of the
method? If yes, please clearly
state the concerns.
Not any apparent from paper
Does the method include all
necessary safety precautions
and/or warning statements? Yes
or No. If No, please suggest
potential precautions or warnings
that can be added to the method
for users or recommend safer
option(s) if warranted.
High potential iTCE source/leads/connectors?
Recommendations
Final Recommendation:
Qualified approval