LOQ:

Not provided

Accuracy/Recovery

Spike recovery: 102.5-103.6% at three spike levels (50, 100

and 150%). The number of repeated spikes is not reported

Precision (RSDr)

The results of 5 sample preparations in two laboratories

were provided. The overall RSDn is 7.11 and that is larger

than required in the SMPR

Reproducibility (RSDR)

System Suitability

NA

Safety Review

Is the method safe to use as

written?

Yes

Are there any safety concerns with

any component, instrument,

reagent or instruction of the

method? If yes, please clearly state

the concerns.

No

Does the method include all

necessary safety precautions

and/or warning statements? Yes or

No. If No, please suggest potential

precautions or warnings that can

be added to the method for users

or recommend safer option(s) if

warranted.

Yes

Recommendations

Final Recommendation:

The method as written is applicable for the analysis of raw

materials and simple finished products . The method

doesn't meet the Method Performance Requirements

stated in the SMPR, but these might be improved after

further method development/optimization. Depending on

the quality of the other submission this method can be a

candidate for further consideration.

CHON-003