Method Submissions-5th Set of Nutrients_6-30-14 - page 192

COMMITTEEDRAFT
ISO/CD 15151 | IDF 229
© ISO2010
All rights reserved
11
AnnexB
(informative)
Noteson thedetection technique, interferences andquantification
B.1 General
AES techniques are widely used for qualitative and quantitative analysis. This annex describes some
phenomena that can be of importance for the interpretation of the procedures of this International Standard.
Although some theoretical considerations are made, this annex is not intended to be a handbook of
spectroscopic techniques.
B.2 Interferences
B.2.1 General
For the determination of a specific analyte in a sample, usually themost sensitive lines are preferred. In case
of interferences, especially spectral interferences, another line has to be selected, even when it is a less
sensitive one. It is known that the ICP-AES technique suffers from a variety of interferences, which are shortly
described hereafter.
B.2.2 Spectral interferences
Spectral line interference, where atomic lines overlap or are unresolved, is often encountered in atomic
emission where light is emitted not only by the element of interest but also from all other elements present in
the sample. Very often this kind of interference can be eliminated by the proper choice of emission line and
can be checked with a second emission line (Table Annex B).
Ensure that the concentration value from the
second check line does not deviate from the first emission line bymore than+/- 5%.
A kind of spectral interference encountered in emission techniques is the occurrence of band-emission
spectradue to the presenceof molecular species.
B.2.3 Ionization interferences
Ionization interferences are caused by the presence of easily ionizable elements in thematrix of the sample,
resulting in a change of the ionization equilibrium of the analyte due to an increase in the electron number
density.
Adding larger amounts of an easily ionizable element to sample and calibration solutions can be used to
overcome this kindof interference.
B.2.4 Physical interferences
Physical interferences are caused by differences in some physical properties of the solutions (sample and
calibration standards) such as viscosity, surface tension and vapour pressure. These differences can then
cause changes inaspiration, nebulization, or atomizationefficiency.
They can be overcome to some extent by applyingmatrix matching of the calibration solutions, by dilution or
by adding relatively highacid concentrations, or bymeans of the standard addition technique.
MTE-03
FORWORKINGGROUP/ERPUSEONLY
DONOTDISTRIBUTE
1...,182,183,184,185,186,187,188,189,190,191 193,194,195,196,197,198,199,200,201,202,...216
Powered by FlippingBook