6. Batch-to-Batch Variation

Measurements of kit standards of several batches were conducted at 450 nm and OD

values were tracked over a 1½ year period.

OD values slightly differed from lot to lot, but the shape of the standard curve stayed the

same.

Figure 1: OD values of standards from various batches measured at 450 nm

Table 2: OD values of standards from various batches measured at 450 nm

Test Date

Standard OD

[450 nm]

0

4

20

80

200

05.01.2015

0,032

0,252

0,745

1,715

2,363

07.01.2015

0,039

0,259

0,738

1,829

2,318

27.01.2015

0,035

0,242

0,708

1,669

2,171

30.01.2015

0,038

0,234

0,678

1,730

2,399

25.02.2015

0,064

0,242

0,825

1,643

2,425

28.04.2015

0,053

0,178

0,606

1,437

2,035

30.04.2015

0,052

0,182

0,628

1,450

2,118

26.06.2015

0,050

0,167

0,507

1,279

1,950

19.08.2015

0,120

0,390

0,999

1,617

2,445

12.09.2015

0,069

0,368

0,875

1,607

2,055

03.01.2016

0,026

0,282

0,711

1,415

1,953

04.02.2015

0,059

0,197

0,508

1,248

1,959

19.02.2015

0,039

0,309

0,948

1,708

2,270

04.03.2016

0,043

0,170

0,530

1,022

1,825

31.03.2016

0,027

0,279

0,889

1,682

2,113

29.04.2016

0,069

0,301

0,921

1,579

2,272

04.05.2016

0,038

0,285

0,770

1,404

1,957

06.05.2016

0,043

0,290

0,853

1,577

2,195

09.05.2016

0,038

0,283

0,854

1,595

2,245

Overall mean, OD

0,049

0,258

0,752

1,537

2,161

Reproducibility SD, OD

0,02

0,06

0,15

0,20

0,19

Reproducibility RSD, %

44

24

20

13

9

0,0

0,5

1,0

1,5

2,0

2,5

3,0

0

20

40

60

80

100 120 140 160 180 200

OD [450 nm]

Kit standards [mg/kg gluten]

Batch-to-Batch Variation of Standards