FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Pros/Strengths:

ICP-AES is a proven, mature technology

MLT run with a great variety of instruments

Variety of microwave digestion systems were used

o

6 CEM MARS

o

3 Milestone Ethos

o

1 Analytikjana TOPWave

o

1 Milestone UltraClave

o

1 NOVAWAVE

o

2 unidentified

Variety of ICP instruments –

o

Old and new models

o

Both Eschelle based optics and Paschen Runge/Roland Circle optics

o

8 Perkin Elmer instruments – 4 models of varying age

o

1 Thermo

o

1 Varian

o

1 Agilent

o

2 Spectro

o

1 Unknown

Both Axial and radial view of plasma used

o

10 labs used axial view for Fe, Cu, Mn and Zn

Great performance on NIST 1849a. No bias for any element.

Not many outliers considering the size of the data set

Cons/Weaknesses

Supporting Data

General Comment:

14 Laboratories from 11 countries

Included SPIFAN matrices plus IDF SAMPLES: milk, milk powder, whey powder, whey protein

concentrate, butter and processed cheese

-

Method Optimization: