Table of Contents Table of Contents
Previous Page  23 / 65 Next Page
Information
Show Menu
Previous Page 23 / 65 Next Page
Page Background

FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Cons

No ISTD

4 point curve used – minimum of 6 levels is recommended for SLV

Method performance requirements:

Lutein

Analytical Range:

Trans Lutein 10-250 µg/L reference standard

6.25 – 156.25 mcg/100g product

Linearity:

0.99991 and <1 % error

LOQ:

0.40 µg/100g RTF

Accuracy/Recovery:

90.3 -95.3 %

Precision (RSD

r

):

1-100 mcg/100g RTF - 1.89 – 7.32 %

14.88 % for child formula?

Intermediated Precision:

1-100 mcg/100g RTF - 1.93 – 8.06 %

13.99 % for Child formula?

Beta Carotene

Analytical Range:

Trans Lutein 25 - 500 µg/L reference standard

15.6 – 312.5 mcg/100g product

Linearity:

0.99993 and <1 % error

LOQ:

0.10 µg/100g RTF

Accuracy/Recovery:

89.3 -108.1 %

Precision (RSD

r

):

1-100 mcg/100g RTF - 1.81 – 6.77 %

Intermediated Precision:

1-100 mcg/100g RTF - 3.07 – 16.20 %

Lycopene

Analytical Range:

Trans Lutein 5-100 µg/L reference standard

3.125 – 62.5 mcg/100g product

Linearity:

0.9998 and <1.5 % error

LOQ:

0.30 µg/100g RTF

Accuracy/Recovery:

97.3 – 108.8 %

Precision (RSD

r

):

1-100 mcg/100g RTF - 3.01 – 6.37 %