FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

3

Performance Characteristics:

Analytical Range:

LOQ:

Accuracy/Recovery

: Great recovery with NIST 1849a. No bias

Reproducibility (RSD

R

): and Precision (RSD

r

):

The overall RSDR for Mn was 12.2%. The SMPR limit is 10%.

The following IF and Adult Nutritional samples did not meet SMPR:

Ca: Adult RTF Hi Fat RSDr=9.3 RSDR=27.8

Infant Elemental Powder RSDr=8.46 RSDR=12.9

Cu: Infant PH Powder Milk RSDr=7.52 Infant Powder Milk RSDSR=12.3

Fe: Adult RTF Hi Fat RSDr=7.81 RSDR=12.8

Adult RTF Hi Protein RSDr=6.14

Infant PH Powder RSDR=14.3

Infant Powder fos-gos RSDR=15.0

Mg: Infant Elemental Powder RSDR=12.1

Mn: Adult RTF Hi Fat RSDr=6.5 RSDR=17.5 Mean = 0.357

Infant PH Powder Milk RSDR20.7

Mean=0.012

Infant Powder fos-gos RSDR=13.0

Mean =0.012

Infant Powder Milk RSDr=6.30 RSDR=23.7 Mean=0.006

Toddler Powder RSDR=11.44

Mean=0.88

IDF Dairy Samples:

Cu: 4 failed both RSDr and RSDR, 1 failed RSDR

Fe: 3 failed botgh RSDr and RSDR, 2 failed RSDR

K: 2 failed RSDR

Mg: 1 failed both RSDr and RSDR

Mn: 2 failed both RSDr and RSDR, 1 Failed RSDR

Zn: 1 failed both RSDr and RSDR, 1 failed RSDR