FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

3

≤ 1.0 mcg/100g RTF 4.67 %

Intermediated Precision:

1-100 mcg/100g RTF - 4.29 – 10.29 %

≤ 1.0 mcg/100g RTF 7.73 %

System suitability:

Correlation of standard curves must be greater than 0.999 and standard response (drift) of not more than

10%. This is a bit high, suggest it should be no more than 3 %. Slope accuracy could be used instead of

standard area to assess drift.

Suggest adding a peak resolution solution as fairly critical for this assay – 5 peaks of interest and depending

on sample matrix, some very close to interfering peaks.

Recommendation:

Recommend to be considered for First Action and MLT Study