FOR EXPERT REVIEW PANEL USE ONLY
NOT FOR DISTRIBUTION
2
•
No major concerns over the chemicals involved
•
Cons/Weaknesses
•
Spike recoveries on the lutein are all below 100%. Is the method under recovering lutein or is the
saponification too harsh on the sample destroying the lutein?
Supporting Data
•
General Comment:
Positive that multiple analysts used in SLV
-
Method Optimization:
Were multiple column manufacturers used for C30
Multiple Instruments?
•
Performance Characteristics:
Analytical Range:
Method meets SMPR requirement
LOQ:
Method Meets SMPR requirement
Accuracy/Recovery:
Recoveries were acceptable except for Lutein. Though they meet the SMPR of 90-110%
requirement, the recoveries are all below 100%.
Precision (RSD
r
):