Table of Contents Table of Contents
Previous Page  465 / 1195 Next Page
Information
Show Menu
Previous Page 465 / 1195 Next Page
Page Background

Manual for details.

1

2

NOTE:

The 3M Molecular Detection Instrument must reach and maintain temperature of 60°C before inserting

3

the 3M Molecular Detection Speed Loader Tray with reaction tubes. This heating step takes approximately 20

4

minutes and is indicated by an ORANGE light on the instrument’s status bar. When the instrument is ready to

5

start a run, the status bar will turn GREEN.

6

7

Analysis

8

Lysis

9

1.

Allow the lysis solution (LS) tubes to warm up to room temperature by setting the rack on the

10

laboratory bench for 2 hours

(a)

.

11

2.

Remove the enrichment broth from the incubator and gently agitate the contents.

12

3.

One LS tube is required for each sample and the Negative Control (NC) sample.

13

3.1

LS tube strips can be cut to desired LS tube number. Select the number of individual LS tubes or

14

8-tube strips needed. Place the LS tubes in an empty rack.

15

3.2

To avoid cross-contamination, decap one LS tubes strip at a time and use a new pipette tip for

16

each transfer step.

17

4.

Transfer enriched sample to LS tubes as described below:

18

19

Transfer each enriched sample into individual LS tube

first

. Transfer the NC

last

.

20

21

4.1

Use the 3M™ Molecular Detection Cap/Decap Tool-Lysis to decap one LS tube strip - one strip at a

22

time. Set the tool with cap attached aside on a clean surface.

23

4.2

Transfer 20 µL of sample into a LS tube.

24

4.3

Repeat step 4.2 until each individual sample has been added to a corresponding LS tube in the

25

strip

26

4.4

Use the 3M Molecular Detection Cap/Decap Tool-Lysis to re-cap the LS tube strip. Use the

27

rounded side of the tool to apply pressure in a back and forth motion ensuring that the cap is

28

tightly applied.

29

30

31

32

4.5

Repeat steps 4.1 to 4.4 as needed, for the number of samples to be tested

33

4.6 When all samples have been transferred, then transfer

20 µL of NC into a LS tube

. Use the 3M

34

Molecular Detection Cap/Decap Tool-Lysis tool to re-cap the LS tube.

35

4.7

Cover the rack of LS tubes with the rack lid and firmly invert 3-5 times to mix.

Suspension has to

36

flow freely inside the tube.

37

5.

Verify that the temperature of the 3M Molecular Detection Heat Block Insert is at 100 ±1°C. Place the

38

rack of LS tubes in the 3MMolecular Detection Heat Block Insert and heat for 15 ±1 minutes

(b)

.

39

Samples that have not been properly heat treated during the assay lysis step may be considered a

40

potential biohazard and should NOT be inserted into the 3M Molecular Detection Instrument.

41

7