AOACRI

Certificate

No.

XXXXXX

Raw shrimp

25 g

225

18-24

Bagged spinach

Pasteurized liquid whole egg

100 g

900

Cooked breaded chicken

325 g

2925

Wet pet food

[Dog – beef cuts in gravy,

canned]

375 g

3375

For other

matrices

Foods

25 g

225

Environmental

1 sponge

225

1

Preparation of the 3M™ Molecular Detection Speed Loader Tray

2

1.

Wet a cloth with a 1-5% (v:v in water) household bleach solution and wipe the 3M™ Molecular

3

Detection Speed Loader Tray.

4

2.

Rinse the 3M Molecular Detection Speed Loader Tray with water.

5

3.

Use a disposable towel to wipe the 3M Molecular Detection Speed Loader Tray dry.

6

4.

Ensure the 3M Molecular Detection Speed Loader Tray is dry before use.

7

8

Preparation of the 3M™ Molecular Detection Chill Block Insert

9

Before using the 3M™ Molecular Detection Chill Block Insert, ensure it has been stored on the 3M™

10

Molecular Detection Chill Block Tray in the freezer (-10 to -20°C) for a minimum of 2 hours before use.

11

When removing the 3M Molecular Detection Chill Block Insert from the freezer for use, remove it and

12

the 3M Molecular Detection Chill Block Tray together. Use the 3M Molecular Detection Chill Block

13

Insert /3M Molecular Detection Chill Block Tray within 20 minutes.

14

15

Preparation of the 3M™ Molecular Detection Heat Block Insert

16

Place the 3M™ Molecular Detection Heat Block Insert in a dry double block heater unit. Turn on the

17

dry block heater unit and set the temperature to 100 ±1°C.

18

19

NOTE:

Depending on the heater unit, allow approximately 30-50 minutes for the 3M Molecular Detection Heat

20

Block Insert to reach temperature. Using a calibrated thermometer, verify that the 3M Molecular Detection Heat

21

Block Insert is at 100 ±1°C.

22

23

Preparation of the 3MMolecular Detection Instrument

24

1.

Launch the 3M™ Molecular Detection Software and log in.

25

2.

Turn on the 3M Molecular Detection Instrument.

26

3.

Create or edit a run with data for each sample. Refer to the 3M Molecular Detection System User

27

6