AOAC EXPERT REVIEW PANEL FOR MICROBIOLOGY FOR FOODS AND ENVIRONMENTAL SURFACES

AOAC Expert Review Panel Final Action Feedback

These factors were addressed in the OMA study considered by the ERP leading to first action OMA

status for the method. The Probability of Detection is < 1 MPN for each of the matrices tested in

validation studies, as evidenced by the fact that none of the >20 matrices tested demonstrated

significantly better performance when comparing the POD for the appropriate reference culture

method(s) and the POD for the BAX® System method. That is, each matrix studied indicated a limit of

detection equal to that of the reference culture method as indicated by a dPOD whose confidence

interval included zero when tested at a fractional level.

ADDITIONAL FEEDBACK FROM USERS OF METHOD

The BAX® System Real Time PCR Assay for Salmonella has had steadily increasing adoptions and sales

since its initial first action status designation indicating robust method performance.