![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0040.png)
AOAC EXPERT REVIEW PANEL FOR MICROBIOLOGY FOR FOODS AND ENVIRONMENTAL SURFACES
AOAC Expert Review Panel Final Action Feedback
These factors were addressed in the OMA study considered by the ERP leading to first action OMA
status for the method. The Probability of Detection is < 1 MPN for each of the matrices tested in
validation studies, as evidenced by the fact that none of the >20 matrices tested demonstrated
significantly better performance when comparing the POD for the appropriate reference culture
method(s) and the POD for the BAX® System method. That is, each matrix studied indicated a limit of
detection equal to that of the reference culture method as indicated by a dPOD whose confidence
interval included zero when tested at a fractional level.
ADDITIONAL FEEDBACK FROM USERS OF METHOD
The BAX® System Real Time PCR Assay for Salmonella has had steadily increasing adoptions and sales
since its initial first action status designation indicating robust method performance.
Page
2
of
13