Table of Contents Table of Contents
Previous Page  116 / 164 Next Page
Information
Show Menu
Previous Page 116 / 164 Next Page
Page Background

Silliker

Chem, Res. Center Crete, IL – Validation of a LC-MS/MS Method for Vitamin K Analysis

Silliker Laboratories, Chemistry Research Center, Date: 1/30/15

10

evaluated range of 5 to 250 ng/mL. Examples of the calibration curve of vitamin K1 peak area and also

peak area as a ratio of peak area of internal std. and also for trans vitamin K1 are provided in Figures

2a, 2b and 2c, respectively. Calibration curve of K2-7 and K2-4 are provided in Figures 3 and 4

respectively.

Figure 2a: One of the example of the Calibration curve (linear regression) for vitamin K1, vitamin K1

conc (ng/mL) (X axis) and peak area for respective standard (YAxis) and their linear regression

relationship.

Figure 2b One of the example of the Calibration curve (linear regression) for vitamin K1, vitamin K1

conc (ng/mL) (X axis) and peak area ratio of vitamin K1 and internal std for respective standard (Y

Axis) and their linear regression relationship.

y = 2233.7x - 1767.6

R² = 0.9995

0

200000

400000

600000

800000

1000000

1200000

0

100

200

300

400

500

600

Vit K1

y = 0.161x - 0.133

R² = 0.9996

0

5

10

15

20

25

30

35

40

45

0

50

100

150

200

250

300

Vit K1 (ratio of int std)

VitK-01 w/SLV

FOR ERP USE ONLY

DO NOT DISTRIBUTE