Table of Contents Table of Contents
Previous Page  448 / 1096 Next Page
Information
Show Menu
Previous Page 448 / 1096 Next Page
Page Background

S433

ESTRO 36

_______________________________________________________________________________________________

Figure 1.

3D printed patient-specific head phantom filled

with dosimetric gel during the treatment planning

process.

Results

Results from mono-energetic irradiation of the cubic

phantoms showed proton range agreement to the TPS

within 1 mm for 90 MeV and 115 MeV, supporting the SPR

gel characterization accuracy. Dose-response linearity was

confirmed for the delivered dose range, except at the

Bragg peak position where a LET dependence was

revealed. Gamma index and relative dose distribution

profiles showed good agreement between TPS and gel, as

shown in in Figure 2.

Figure 2.

(A) Slice of the 3D SFUD dose distribution

converted from a T2 relaxation map obtained from MR

scanning

the irradiated 3D printed head phantom filled with

polymer gel. The PTV is indicated in white. (B) Gel

(RTsafe) and

TPS

dose profiles along the path marked in red in (A). (C)

3%/2mm gamma index along the profile.

Conclusion

In this work we have shown that patient-specific 3D

polymer gel dosimetry is applicable to PT using PBS.

Further characterization and correction of the LET

dependence and comparison to MC dose calculations will

be carried out and presented.

Acknowledgements:

DFG-MAP

PO-0810 Absolute dose pre-treatment Portal Dosimetry

using the Varian MAASTRO implementation

A. Taborda

1

, J. Stroom

1

, C. Baltes

2

, A. Seabra

1

, K.

Dikaiou

2

, C. Greco

1

1

Champalimaud Centre for the Unknown, Clinical

Department, Lisboa, Portugal

2

Varian Medical Systems, Varian Medical Systems Imaging

Laboratory, Baden-Dättwil, Switzerland

Purpose or Objective

Current clinical portal dosimetry from Varian evaluates

dose using calibrated units (CU). This work assesses the

quality of the first Varian implementation of the MAASTRO

algorithm for pre-treatment absolute portal dosimetry (in

Gy) of 6X-FF fields.

Material and Methods

To achieve the proposed goal, a comparison was made

between the gamma analysis results obtained using both

Varian’s clinical portal dosimetry (PDIP v10.0.28) and the

MAASTRO algorithm [1] implementation made available

for the authors in the Portal Dosimetry (PD) application

accessible through the Varian Citrix Research Environment

(CRE). For this study, 10 breast IMRT breast plans and 10

VMAT prostate plans were chosen from the patients’

database. In total, 71 IMRT fields and 40 VMAT arcs were

compiled for analysis. Each plan was recalculated with

gantry zero on a water-equivalent slab phantom, for later

comparison of absolute dose at 5cm depth. Verification

plans were created for irradiation with 6X-FF beams at the

Varian Edge LINAC in order to measure the doses at the

Electronic Portal Imaging Device (EPID) level. For each

field/arc, the measured doses and the calculated doses

were compared by gamma analysis in CU for PDIP and in

absolute dose values (Gy) for the PD system on the CRE.

[1] Nijsten SM et al, 'A global calibration model for a-Si

EPIDs used for transit dosimetry”, Med. Phys. 34(10):

3872-84, 2007

Results

Table 1 presents the summary of the gamma analysis

results obtained in the comparison between the measured

dose at the EPID and the calculated dose using the

MAASTRO algorithm implementation and PDIP. The results

show that the analyzed IMRT plans using the MAASTRO

algorithm obtained, on average, a higher gamma pass

rate, lower mean gamma values and lower dose

differences than while using PDIP. The same is observed

for VMAT plans. Figure 1 shows the graphical comparison

between the gamma passing rate obtained using the

MAASTRO algorithm and PDIP, where the black circles

represent the comparison of the gamma passing rates for

IMRT plans (averaged over all beams) and the open

triangles represent the comparison of the gamma passing

rates for VMAT plans (averaged over all arcs). One can see

that the gamma pass rate obtained using the MAASTRO

algorithm is consistently higher than the one obtained

using PDIP.

Table

1