Table of Contents Table of Contents
Previous Page  499 / 1096 Next Page
Information
Show Menu
Previous Page 499 / 1096 Next Page
Page Background

S484

ESTRO 36

_______________________________________________________________________________________________

Results

Figure 2 illustrates that proxy-free and common external

breathing signal-driven PB-reconstructed 4DCT data are

comparable both in terms of image quality and

represented motion amount. In detail, the considered

proxy-free datasets contained approximately 5% more

artifacts than the PB data sets. Differences of represented

tumor mass center motion as well as the amount of e.g.

diaphragm motion between end-inhalation and -

exhalation were negligible (max. 1 voxel).

Conclusion

We presented a novel approach for proxy-free slow-pitch

helical 4DCT reconstruction and illustrated its feasibility.

Although the proxy-free reconstructed images contain

slightly more motion artifacts, we consider the approach

to be helpful especially in the case of corrupted breathing

signals recordings (no need for re-scanning the patient).

PO-0883 Clinical Implementation Model-Based CT to

Replace 4DCT for Lung Cancer Treatment Planning

D. Low

1

, D. O'Connell

1

, L. Yang

1

, J. Lewis

1

, P. Lee

1

1

UCLA Medical Center, Department of Medical Physics, Los

Angeles, USA

Purpose or Objective

To implement motion-model based CT into clinical

practice, replacing 4DCT for breathing motion

management treatment planning.

Material and Methods

A breathing motion model that employs a mathematical

motion equation, two real-time breathing surrogates,

breathing amplitude and breathing rate, and employing

multiple fast helical, low-dose CT scanning has been

introduced into clinical practice. The imaging process uses

a bellows-based system to monitor the breathin g cycle,

which is defined as the amplitude and rate of the bellows

signal. The fast helical CT scans are reg istered to

determine the lung tissue positions, correlate d to the

breathing amplitude and rate on a slice-by-s lice basis. A

published motion equation is employed to characterize

the motion for each voxel. The motion model is employed

to reconstruct the original fast helical CT scans and the

original and reconstructed scans compared to determine

the overall model motion prediction accuracy. Eight

amplitude-based CT images are constructed and sent to

the treatment planning system, along with a three-

dimensional motion model accuracy (defined as the 75th

percentile motion error in each voxel) map. The patients

still undergo a commercial 4DCT protocol to provide a

comparison between the current standard of care and the

model-based process. Comparisons between the

commercial and model-based approaches have been

conducted on 19 patients to evaluate the magnitude of

sorting artifacts in each process on a scale of 1-4, 1 having

no artifacts and 4 having severe artifacts. The average CT

noise for both protocols was described by examining a

region of interest in the liver.

Results

Mean tumor displacement was 11.5 +/- 6.9 mm and the

mean motion model error was 1.77 +/- 0.79 mm. The mean

artifact severity ratings for the 4DCT and model-based CT

approaches were 2.2 and 1.2, respectively. There were

three instances of grade 4 artifacts and no instances of

grade 3 or worse artifacts for the 4D and model-based

approaches, respectively. The average CT noise was

reduced from 57.7 HU to 11.6 HU.

Conclusion

The model-based approach provides the clinic with motion

artifact free images that have lower noise and whose

geometry accurately reflects the tumor and other lung

tissues during the CT scanning session. We are still limited

by the treatment planning system's input requirements for

a series of breathing-phase defined images. Work is

ongoing to develop treatment planning protocols that

better match the data resulting from the model-based

approach.

PO-0884 Availability of MRI improves interobserver

variation in CT-based pancreatic tumor delineation

O.J. Gurney-Champion

1

, E. Versteijne

1

, A. Van der

Horst

1

, E. Lens

1

, H. Rütten

2

, H.D. Heerkens

3

, G.M.R.M.

Paardekooper

4

, M. Berbee

5

, C.R.N. Rasch

1

, J. Stoker

6

,

M.R.W. Engelbrecht

6

, M. Van Herk

7

, A.J. Nederveen

6

, R.

Klaassen

8

, H.W.M. Van Laarhoven

8

, G. Van Tienhoven

1

, A.

Bel

1

1

Academic Medical Center, Department of Radiation

Oncology, Amsterdam, The Netherlands

2

Radboud University Medical Center, Department of

Radiation Oncology, Nijmegen, The Netherlands

3

University Medical Center Utrecht, Department of

Radiotherapy, Utrecht, The Netherlands

4

Isala Clinics Zwolle, Department of Radiotherapy,

Zwolle, The Netherlands

5

MAASTRO Clinic, Department of Radiation Oncology,

Maastricht, The Netherlands

6

Academic Medical Center, Department of Radiology,

Amsterdam, The Netherlands

7

University of Manchester and Christie NHS trust, Faculty

of Biology- Medicine & Health- Division of Molecular &

Clinical Cancer Sciences, Manchester, United Kingdom

8

Academic Medical Center, Department of Medical

Oncology, Amsterdam, The Netherlands

Purpose or Objective

To assess whether the availability of magnetic resonance

images (MRIs) alongside the planning CT scan for target

volume delineation in pancreatic cancer patients

decreases interobserver variation.

Material and Methods

Eight observers (radiation oncologists) from six institutions

delineated gross tumor volume (GTV) on contrast-

enhanced (CE) 3DCT and internal GTV (iGTV) on 4DCT for

four pancreatic cancer patients. At least six weeks after

submitting these delineations, the observers were asked

to repeat the delineations, now with MRIs available in a

separate window (3DCT+MRI and 4DCT+MRI). The MRI

included plain and CE T1-weighted gradient echo, T2-

weighted turbo spin echo, and diffusion-weighted

imaging. Interobserver variation in volumes of (i)GTVs was

analyzed. Also, the generalized conformity index (CI

gen

), a