Table of Contents Table of Contents
Previous Page  520 / 1096 Next Page
Information
Show Menu
Previous Page 520 / 1096 Next Page
Page Background

S505

ESTRO 36

_______________________________________________________________________________________________

average dose, following irradiation with periodically

modulated and open 6MV photon fields.

Results

Survival of normal cells in a 2.5mm striped field was the

same as for an open field, but the survival of the cancer

cells was significantly lower. However for cancer cell lines

in the 5.0mm modulated fields, the response compared to

an open field was no longer statistically significant. A

mathematical model was developed to incorporate the

dose gradients of the spatial modulation into the standard

linear quadratic model. Our new extended bystander LQ

model assumes spatial gradients drive the diffusion of

soluble factors that influence survival through bystander

effects. The model successfully predicts the experimental

results that show an increased therapeutic benefit.

Conclusion

We have confirmed that HDMLCs can create spatially

modulated fields that increase the therapeutic advantage

between normal and cancer cells. Our results challenge

conventional radiotherapy practice and propose that

additional gain can be realized by prescribing spatially

modulated treatments to harness the bystander effect.

PO-0912 Short- and long term stability of the isocenter

of a three-source Co60 MR guided radiotherapy device

D. Hoffmans

1

, M.A. Palacios

1

, J.P. Cuijpers

1

1

VU University Medical Center, Radiotherapy,

Amsterdam, The Netherlands

Purpose or Objective

Recently a 0.35T Co

60

MRIdian system (Viewray Inc.,

Cleveland) is implemented at our institution. In a similar

way as for other image guided radiotherapy techniques,

the coincidence of the radiation therapy (RT) and imaging

isocenter is of major importance. The purpose of this

study is to present a method for daily QA of MR-RT

isocenter coincidence and to assess its short- and long

term stability using daily film-based isocenter QA.

Material and Methods

Two pieces of radiochromic film (GafChromic EBT3) are

taped to square inlays on the top and the side of a cubic

water-filled phantom. The phantom is aligned to the MRI

isocenter using MR guided setup relying on three internal

cylindrical markers.

A treatment plan, consisting of an AP and a lateral square

field is delivered. The direction of the lateral field is

altered daily between 90

o

and 270

o

in order to monitor all

3 treatment heads (Head 1 and 2 for the lateral fields

respectively, head 3 for the AP field). The films are

digitized and the positions of the square fields with

respect to the phantom are determined. These data

provide a daily measurement of the coincidence of the RT-

and MR- isocenters in 3 dimensions. The AP field provides

information about the alignment along the lateral (X) and

the longitudinal (Y) axis. The lateral fields provide the

shift in the vertical (Z), and the Y directions. Data were

collected daily over a period of 4 months. A linear

regression is performed in order to determine any trends

in time. Furthermore, the correlation between the two

daily values for the shift along the Y-axis is assessed.

Results

The average 3D vector of the daily shift is found to be

0.8mm (P95 = 1.3 mm). The average misalignments as

determined by the individual heads are shown in Table 1.

The systematic shift in 3D is zero in X and Y direction and

0.2 mm in the Z direction, which is caused by th e vertical

shift measured with head 1. No time trend in the shift is

observed in any direction as the regression coefficients

were not statistically significant different from zero:

p=0.39, 0.64 and 0.50 for the X, Y and Z axes respectively.

The pearsman correlation coefficient between the Y-shift

determined using the two perpendicular fields was very

weak and found to be 0.24.

Conclusion

A method is developed for daily assessment of the

coincidence of the MV- and MR-isocenter for an integrated

MR-RT unit. The alignment of the MR- to the RT-isocenter

is found to be stable during a time period of 4 months. A

small systematic shift in vertical direction was found, a

star shot measurement confirmed that this was caused by

a slight misalignment of Head 1. This misalignment can be

compensated by realignment of the MLC leaf positions.

The weak correlation in the pair of Y-measurements

suggests that the daily misalignment is dominated by

random measurement inaccuracies such as the placing

markers on the film and rotational setup misalignments of

the phantom.

PO-0913 A national review of equipment, techniques

and PTV margins used for SRS

R. Patel

1

, J. Lee

1

, C. Walker

2

, D.J. Eaton

1

1

Mount Vernon Hospital, Radiotherapy RTTQA,

Northwood Middlesex, United Kingdom

2

Northern Centre for Cancer Care, Radiotherapy,

Newcastle, United Kingdom

Purpose or Objective

As part of a national commissioning programme,

treatment providers were required to complete a SRS

quality assurance review in order to benchmark current

practice. The process was designed to ensure providers

were able to deliver a service in line with parameters

developed by a multidisciplinary expert advisory group

(EAG).

The long term aim of this programme was to progress a

system of standardisation and quality improvement of

service by promoting consistency and the development of

services over time. The short term goals were to highlight

any significant variation in practice in order to identify

centres that may require further support or mentoring in

order to meet nationally agreed parameters.

Material and Methods

A questionnaire was circulated to 20 centres to establish

the equipment, treatment techniques and PTV expansions

used to deliver SRS. Centres reported on their current