S507
ESTRO 36
_______________________________________________________________________________________________
in-line MV EPID. The MPC acquires a series of MV images
of an IsoCal phantom, capturing beam properties and
mechanical data such as MLC and gantry accuracies. A
new MPC test monitors output stability in terms of percent
deviation from the baseline data of the actual measured
beam. All measured data are automatically processed,
analyzed, and displayed for evaluation, thus providing a
reliable and fast method for routine machine performance
assessment. Independent tests such as star-shots,
Winston-Lutz, MLC picket fence patterns and output
measurements on a daily basis were employed to
benchmark the MPC test results for the prototype system.
Results
MPC results were collected daily for six months on both
the prototype and a TB. The independent tests on the
prototype system were repeated weekly to validate the
MPC results. A sample comparison of the MPC results for
the prototype against independent tests are shown in
Table 1. The output stability of the prototype system, as
measured with the MPC and a DailyQA
TM
3 device, is
comparable (Fig. 1), and within 0.5% of independent
output measurement for the period shown. All tests
performed were within the tolerances allowed by the MPC
and agreed in most cases with the result of the
independent tests. The prototype system performs as well
as the TB system. A summary of MPC test results and
comparisons with independent measurements will be
shown alongside with the TB MPC results.
Table 1
Figure 1
Conclusion
The performance of a straight-through linac d elivering 6X-
FFF (filter-free) beam with an EPID panel wa s investigated
with the MPC testing tool and that method w as validated
against independent tests for proof of con cept. MPC is a
complete, reliable and quick test suite that monitors the
performance of a treatment unit on routine basis.
PO-0916 Feasibility and potential for treating loca lly
advanced non-sma ll cell lung cancer with a MR-linac
M.J. Menten
1
, H. Bainbridge
2
, M.F. Fast
1
, S . Nill
1
, F.
McDonald
2
, U. Oelfke
1
1
The Institute of Cancer Research and The Royal Marsden
NHS Foundation Trust, Physics, Sutton, United Kingdom
2
The Institute of Cancer Research and The Royal Marsden
NHS Foundation Trust, Lung Unit, Sutton, United
Kingdom
Purpose or Objective
Treatment plans for MRI-guided radiotherapy delivered
with an MR-linac vary from those designed for
conventional linacs due to differing technical
specifications of dose delivering systems and the presence
of a static magnetic field. This study investigated this
issue for radiotherapy of locally advanced non-small cell
lung cancer (LA NSCLC) by comparing treatment plans for
a conventional Versa HD linac (Elekta AB, Stockholm,
Sweden) and the Elekta 1.5 T MR-linac. Furthermore, the
effect of reducing planning target volume (PTV) margins
on the MR-linac was examined.
Material and Methods
Ten patients with LA NSCLC were retrospectively re-
planned six times using the Monaco treatment planning
system, research version 5.19.00. Three plans were
designed according to our institution’s protocol for
conventionally fractionated treatment (55 Gy/ 20
fractions) and three plans following guidelines for isotoxic
dose escalation up to 79.2 Gy/ 44 fractions
(NCT01836692). In each case, two plans were designed for
the MR-linac, using IMRT with nine equidistant, coplanar
beams, either with standard (7 mm) or reduced (3 mm)
PTV margins, while one plan was created for a
conventional linac using VMAT with standard margins.
Treatment plan optimization and dose calculation were
conducted under consideration of magnetic field effects.
Potential to escalate tumour dose was quantified for the
isotoxic plans, and differences in dose-volume metrics
were analysed for conventionally fractionated treatment
plans. Statistical significance was evaluated using a
paired
t-
test after confirming normal distribution and
correcting for multiple endpoints.
Results
All generated treatment plans fulfilled their respective
planning constraints and would have been clinically
acceptable. With the conventionally fractionated
schedule small differences in dose-volume metrics could
be identified with statistical significance (see table). Mean
lung doses were similar between conventional and MR-
linac plans, whereas high lung doses were reduced and low
lung doses increased on the MR-linac (graphically
illustrated in the figure). In terms of dose-escalation, the
mean achievable doses were 75.4, 74.0, and 76.9 Gy for
Versa HD, MR-linac (standard margins) and MR-linac
(reduced margins) respectively, with inferiority of the
standard margin MR-linac plans versus the Versa HD plans
(
p
=0.003).