Table of Contents Table of Contents
Previous Page  525 / 1096 Next Page
Information
Show Menu
Previous Page 525 / 1096 Next Page
Page Background

S510

ESTRO 36

_______________________________________________________________________________________________

beams (0.4 ± 3.2%). Comparing C4 to C2, log file accuracy

is analyzed for MC. LF resulted in lower σ values for 20/22

arcs (-5.4 ± 3.4%) and improved pass rates for 14/18 arcs

(1.1 ± 1.4%). Comparing C5 to C2, LF and AC QA

techniques are compared. The LF technique yielded

decreased σ values for 22/22 arcs (-51 ± 7%) and improved

pass rates for 18/18 fields (9.9 ± 3.8%). The LF technique

also eliminated systematic AC errors; mean dose errors

decreased from 3.2% to 0.1%. For 1/2/3/4° LF-CS control

point spacing, 1%/1mm pass rates were 80.0 ± 5.0%, 78.0

± 4.2%, 74.0 ± 5.1%, and 68.8 ± 5.3%. Plan-CS pass rates

were 80.2 ± 4.0%. Figure 2 plots difference in pass rates

[(LF-CS vs. AC) minus (Plan-CS vs. AC)] as a function of

control point spacing for each arc. Calculation times for

CS and MC were 12s per control point and 3 minutes per

VMAT arc respectively.

Conclusion

MC doses proved more accurate than CS when compared

to AC measurement. LF-MC plans yielded superior

accuracy and shorter calculation times than LF-CS plans.

By cutting out the phantom and comparing LF dose to that

of the original plan, systematic error was eliminated and

random error greatly reduced.

PO-0921 Dose considerations of IGRT using MV

projection and MV CBCT on a prototype linear

accelerator

P. Balter

1

, T. Netherton

1

, Y. Li

1

, P. Nitsch

1

, S. Gao

1

, M.

Muruganandham

1

, S. Shaitelman-

1

, S. Frank

1

, S. Hahn

1

, A.

Klopp

1

, L. Court

1

1

UT MD Anderson Cancer Center Radiation Physics,

Radiation Physics, Houston- TX, USA

Purpose or Objective

The use of the mega-voltage treatment beam for image-

guided patient setup has some potential advantages over

kV imaging, especially reduced equipment and QA

requirements. One of the challenges that MV imaging

introduces is the increase in daily imaging dose. Here we

investigate (1) whether the MV imaging dose can be

correctly calculated and incorporated into the treatment

plan, and (2) the impact of MV imaging dose on the dose

to normal tissues such as the lung and heart.

Material and Methods

MV imaging dose to the lung, heart and other soft tissue

was measured using an ion chamber in anthropomorphic

thorax phantom (CIRS), and compared with dose

calculated in the TPS (Eclipse) for orthogonal MV-MV

imaging fields and MV CBCT images using a prototype

linear accelerator, each with a low-dose and high-quality

mode (total 4 modes). The impact of the imaging

technique (orthogonal vs. CBCT and high vs. low quality)

on the doses to normal tissue was evaluated using Eclipse,

where the imaging doses were used as based plans in the

treatment planning process. For breast plans, doses to the

heart and lung were evaluated. For head/neck plans,

doses to all the normal tissues were compared.

Figure 1. Anthropomorphic phantom (CIRS) with dose

measurement points identified.

Results

Average imaging dose was measured as 1.3, 2.5, 3.7, and

7.6cGy for daily low dose MV pairs, high quality MV pairs,

low dose CBCT and high quality CBCT, respectively. Over

a 30 fraction treatment with daily IGRT, this equates to 38

- 227cGy. The average agreement between measured and

calculated tissue doses due to imaging was 0.4±0.4cGy.

The largest difference was 1.3cGy, found in the lung for

high quality CBCT imaging (~39cGy over a 30 fraction

treatment).

With imaging dose incorporated into the treatment

planning process, it was possible to create clinically

acceptable treatment plans for a range of treatment sites,

including breast, head and neck and prostate. The imaging

technique did, however, increase the heart and lung dose

for breast plans. For an example left breast treatment,

the mean heart dose in our original, clinically delivered

plan was 60cGy. With daily MV imaging included, this

increased to 140, 150, 190 and 260cGy for daily low dose

MV pairs, high quality MV pairs, low dose CBCT and high

quality CBCT, respectively. The corresponding values for

mean lung dose were 360cGy (original clinical) and 470,

490, 510 and 570cGy.

Table 1: Tissue doses(cGy/fraction) at different points in

the anthropomorphic phantom. M: Measured. C:

Calculated

Conclusion