Ruggedness
4.6.1) Definition:
Ruggedness of an analytical method is a measure of degree of reproducibility
of results obtained under different environmental conditions such as different laboratory,
different chemist and different instrument.
Purpose of ensuring the method under validation provides the same results when analysed
in a different HPLC, different brand column and different conditions.
4.6.2) Assessment:
Withanoside IV, Withanoside V, Withaferin A, 12-Doxy withastramonolide,
Withanolide A and Withanolide B
contents in extracts sample and raw material were
analysed by two different analysts at different dates using different instruments and C18
column from a different manufacturer. The results were compared and RSD was
calculated.
Table-64:
Name of the instrument
Condition–1
Condition-2
HPLC
LC-2010A
LC10ATvp
C18 Column
Lichrospher 100
RP-18e(5
µ
m)
Col. No. 625364
Make: Merck (Hiber)
Phenomenex, Luna
5 µ C
18
(2)100 A,
250 X 4.60 mm, 5 microns
Part No. :00G-4252- E0
No. 359208-12
Column oven temperature
25°C
30°C
Concentration of H
3
PO
4
0.5ml in 1000ml
1.5ml in 1000ml
4.6.3) Acceptance criteria:
The RSD between the two results obtained two different analysts
using two different instruments on different date should be less than 6.0%.
4.6.4) Experiment:
Experiment-1 (Condition - 1)
a) Standard / sample preparation: Refer 4.3.11
Table-65:
Batch No.
Withanoside
IV
Withanoside
V
Withaferin
A
12-Deoxy
withastra
monolide
Withanolide
A
Withanolide
B
Sum
WS/06Lot08
0.917
0.650
0.659
0.284
0.225
0.066
2.802
WS/06Lot10
0.911
0.621
0.616
0.267
0.205
0.062
2.681
WS/05Lot20
0.067
0.044
0.081
0.022
0.021
0.005
0.241
WS/05Lot21
0.478
0.310
0.410
0.125
0.102
0.029
1.453
RD/1170
0.848
0.594
0.700
0.332
0.229
0.076
2.779
RD/1045
0.606
0.364
0.402
0.248
0.131
0.042
1.793
RD/1162
0.152
0.090
0.100
0.021
0.025
0.004
0.391
ERH-46
0.184
0.148
0.145
0.041
0.041
0.010
0.570